[Nellix® endovascular aneurysm sealing (EVAS) - a new technology for endovascular management of infrarenal aortic aneurysms]. / Nellix® Endovascular Aneurysm Sealing (EVAS) - eine neue Technologie zur endovaskulären Ausschaltung infrarenaler Aortenaneurysmen.
Böckler, D; Peters, A S; Pfeiffer, S; Kovacs, B; Geisbüsch, P; Bischoff, M S; Müller-Eschner, M; Hakimi, M; Pfeiffer, S.
Afiliación
  • Böckler D; Klinik für Gefäßchirurgie und Endovasculäre Chirurgie, Universitätsklinik Heidelberg, Deutschland.
  • Peters AS; Klinik für Gefäßchirurgie und Endovasculäre Chirurgie, Universitätsklinik Heidelberg, Deutschland.
  • Pfeiffer S; Klinik für Gefäßchirurgie und Endovasculäre Chirurgie, Universitätsklinik Heidelberg, Deutschland.
  • Kovacs B; Klinik für Gefäßchirurgie und Endovasculäre Chirurgie, Universitätsklinik Heidelberg, Deutschland.
  • Geisbüsch P; Klinik für Gefäßchirurgie und Endovasculäre Chirurgie, Universitätsklinik Heidelberg, Deutschland.
  • Bischoff MS; Klinik für Gefäßchirurgie und Endovasculäre Chirurgie, Universitätsklinik Heidelberg, Deutschland.
  • Müller-Eschner M; Klinik für Diagnostische und Interventionelle Radiologie, Universitätsklinik, Heidelberg, Deutschland.
  • Pfeiffer S; Klinik für Gefäßchirurgie und Endovasculäre Chirurgie, Universitätsklinik Heidelberg, Deutschland.
Zentralbl Chir ; 139(5): 562-8, 2014 Oct.
Article en De | MEDLINE | ID: mdl-25313891
ABSTRACT

BACKGROUND:

Recently used endografts for envascular aneurysm repair (EVAR) exclude the pathology by fixation at both the proximal and distal landing zone. Due to endoleaks and migration EVAR is associated with a relevant rate of secondary interventions. The Nellix® system (Endologix Inc., CA, USA) was developed to seal the complete aneurysm using a polymer filling, therefore stabilising endograft-position and reducing the rate of endoleaks and reinterventions. The present contribution introduces the method, describes the technique of implantation and presents the first clinical results. Material und

Methods:

The Nellix system consists of two balloon-expandable stent grafts made of a cobalt-chromium composition, surrounded with ePTFE and the so-called endobags. During the implantation each endobag is filled with a non-biodegradable polymer, sealing the aneurysm lumina including the proximal and distal landing zone. Hence, lumbar arteries will be sealed to reduce the probability of a type II endoleak.

RESULTS:

Longterm durability as well as the structural integrity of the Nellix system has been proven over 4 years in sheep experiments. The technical success in a multicentre, prospective registry was 94% without the appearance of severe adverse events (migration, occlusion, secondary endoleak).

CONCLUSION:

EVAS is a new and different concept of endovascular AAA repair. Recent clinical data of the Nellix system are promising showing a high technical success rate while the need for secondary intervention is low. Further studies in larger cohorts are needed.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Complicaciones Posoperatorias / Diseño de Prótesis / Prótesis Vascular / Stents / Aneurisma de la Aorta Abdominal / Endofuga / Procedimientos Endovasculares Límite: Animals / Humans Idioma: De Revista: Zentralbl Chir Año: 2014 Tipo del documento: Article

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Complicaciones Posoperatorias / Diseño de Prótesis / Prótesis Vascular / Stents / Aneurisma de la Aorta Abdominal / Endofuga / Procedimientos Endovasculares Límite: Animals / Humans Idioma: De Revista: Zentralbl Chir Año: 2014 Tipo del documento: Article