Hohlraum glint and laser pre-pulse detector for NIF experiments using velocity interferometer system for any reflector.
Moody, J D; Clancy, T J; Frieders, G; Celliers, P M; Ralph, J; Turnbull, D P.
Afiliación
  • Moody JD; Lawrence Livermore National Laboratory, Livermore, California 94550, USA.
  • Clancy TJ; Lawrence Livermore National Laboratory, Livermore, California 94550, USA.
  • Frieders G; Lawrence Livermore National Laboratory, Livermore, California 94550, USA.
  • Celliers PM; Lawrence Livermore National Laboratory, Livermore, California 94550, USA.
  • Ralph J; Lawrence Livermore National Laboratory, Livermore, California 94550, USA.
  • Turnbull DP; Lawrence Livermore National Laboratory, Livermore, California 94550, USA.
Rev Sci Instrum ; 85(11): 11E608, 2014 Nov.
Article en En | MEDLINE | ID: mdl-25430354
ABSTRACT
Laser pre-pulse and early-time laser reflection from the hohlraum wall onto the capsule (termed "glint") can cause capsule imprint and unwanted early-time shocks on indirect drive implosion experiments. In a minor modification to the existing velocity interferometer system for any reflector diagnostic on NIF a fast-response vacuum photodiode was added to detect this light. The measurements show evidence of laser pre-pulse and possible light reflection off the hohlraum wall and onto the capsule.

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Idioma: En Revista: Rev Sci Instrum Año: 2014 Tipo del documento: Article País de afiliación: Estados Unidos