NT-proBNP on Cobas h 232 in point-of-care testing: Performance in the primary health care versus in the hospital laboratory.
Gils, Charlotte; Ramanathan, Ramshanker; Breindahl, Torben; Brokner, Mette; Christiansen, Anne L; Eng, Øyvin; Hammer, Ida J; Herrera, Catherine B; Jansen, Aina; Langsjøen, Eva C; Løkkebo, Elisabeth S; Osestad, Toril; Schrøder, Anne D; Walther, Lisa.
Afiliación
  • Gils C; a Department of Clinical Biochemistry and Pharmacology , Odense University Hospital , Odense , Denmark.
Scand J Clin Lab Invest ; 75(7): 602-9, 2015 Nov.
Article en En | MEDLINE | ID: mdl-26305423
ABSTRACT

BACKGROUND:

NT-proBNP may be useful for ruling out heart failure in primary health care. In this study we examined the analytical quality of NT-proBNP in primary health care on the Cobas h 232 point-of-care instrument compared with measurements performed in a hospital laboratory. MATERIALS AND

METHODS:

Blood samples requested for NT-proBNP were collected in primary health care (n = 95) and in a hospital laboratory (n = 107). NT-proBNP was measured on-site on Cobas h 232 instruments both in primary health care centres and at the hospital laboratory and all samples were also analyzed with a comparison method at the hospital. Precision, trueness, accuracy, and lot-variation were determined at different concentration levels and evaluated according to acceptance criteria. Furthermore user-friendliness was assessed by questionnaires.

RESULTS:

For Cobas h 232 repeatability CV was 8.5-10.7% in the hospital setting and 5.3-10.0% in the primary health care and within the analytical quality specifications, but higher than with the comparison method (< 4%). NT-proBNP results obtained in primary health care were significantly higher than by the hospital comparison method (bias ranged from 14.3-23.7%), whereas there was no significant bias when Cobas h 232 was used in the hospital setting (bias ranged from - 4.9 to 7.0%). User-friendliness of Cobas h 232 was overall acceptable.

CONCLUSION:

Cobas h 232 point-of-care instrument for measurement of NT-proBNP performed satisfactorily with regard to precision, user-friendliness, and lot-variation. A decrease in NT-proBNP levels observed in samples transported to a central laboratory needs further attention and investigation.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Fragmentos de Péptidos / Atención Primaria de Salud / Laboratorios de Hospital / Péptido Natriurético Encefálico / Pruebas en el Punto de Atención Tipo de estudio: Diagnostic_studies / Evaluation_studies Límite: Humans Idioma: En Revista: Scand J Clin Lab Invest Año: 2015 Tipo del documento: Article País de afiliación: Dinamarca Pais de publicación: ENGLAND / ESCOCIA / GB / GREAT BRITAIN / INGLATERRA / REINO UNIDO / SCOTLAND / UK / UNITED KINGDOM

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Fragmentos de Péptidos / Atención Primaria de Salud / Laboratorios de Hospital / Péptido Natriurético Encefálico / Pruebas en el Punto de Atención Tipo de estudio: Diagnostic_studies / Evaluation_studies Límite: Humans Idioma: En Revista: Scand J Clin Lab Invest Año: 2015 Tipo del documento: Article País de afiliación: Dinamarca Pais de publicación: ENGLAND / ESCOCIA / GB / GREAT BRITAIN / INGLATERRA / REINO UNIDO / SCOTLAND / UK / UNITED KINGDOM