Ion-selective electrodes in potentiometric titrations; a new method for processing and evaluating titration data.
Granholm, Kim; Sokalski, Tomasz; Lewenstam, Andrzej; Ivaska, Ari.
Afiliación
  • Granholm K; Process Chemistry Centre, c/o Laboratory of Analytical Chemistry, Åbo Akademi University, Biskopsgatan 8, FIN-20500 Åbo-Turku, Finland. Electronic address: kim.granholm@abo.fi.
  • Sokalski T; Process Chemistry Centre, c/o Laboratory of Analytical Chemistry, Åbo Akademi University, Biskopsgatan 8, FIN-20500 Åbo-Turku, Finland.
  • Lewenstam A; Process Chemistry Centre, c/o Laboratory of Analytical Chemistry, Åbo Akademi University, Biskopsgatan 8, FIN-20500 Åbo-Turku, Finland; AGH - University of Science and Technology, Faculty of Materials Science and Ceramics, Mickiewicza 30, 30-059 Krakow, Poland.
  • Ivaska A; Process Chemistry Centre, c/o Laboratory of Analytical Chemistry, Åbo Akademi University, Biskopsgatan 8, FIN-20500 Åbo-Turku, Finland.
Anal Chim Acta ; 888: 36-43, 2015 Aug 12.
Article en En | MEDLINE | ID: mdl-26320956
A new method to convert the potential of an ion-selective electrode to concentration or activity in potentiometric titration is proposed. The advantage of this method is that the electrode standard potential and the slope of the calibration curve do not have to be known. Instead two activities on the titration curve have to be estimated e.g. the starting activity before the titration begins and the activity at the end of the titration in the presence of large excess of titrant. This new method is beneficial when the analyte is in a complexed matrix or in a harsh environment which affects the properties of the electrode and the traditional calibration procedure with standard solutions cannot be used. The new method was implemented both in a method of linearization based on the Grans's plot and in determination of the stability constant of a complex and the concentration of the complexing ligand in the sample. The new method gave accurate results when using titrations data from experiments with samples of known composition and with real industrial harsh black liquor sample. A complexometric titration model was also developed.