Your browser doesn't support javascript.
loading
Optimal Vagus Nerve Stimulation Frequency for Suppression of Spike-and-Wave Seizures in Rats.
Jiao, Jianhang; Harreby, Kristian R; Sevcencu, Cristian; Jensen, Winnie.
Afiliación
  • Jiao J; Center for Sensory-Motor Interaction, Department of Health Science and Technology, Faculty of Medicine, Aalborg University, Aalborg, Denmark.
  • Harreby KR; Center for Sensory-Motor Interaction, Department of Health Science and Technology, Faculty of Medicine, Aalborg University, Aalborg, Denmark.
  • Sevcencu C; Center for Sensory-Motor Interaction, Department of Health Science and Technology, Faculty of Medicine, Aalborg University, Aalborg, Denmark.
  • Jensen W; Center for Sensory-Motor Interaction, Department of Health Science and Technology, Faculty of Medicine, Aalborg University, Aalborg, Denmark.
Artif Organs ; 40(6): E120-7, 2016 Jun.
Article en En | MEDLINE | ID: mdl-26713661
ABSTRACT
Vagus nerve stimulation (VNS) is used as an adjunctive therapy for drug-resistant epilepsy and results in a 50% seizure reduction in up to 50% of treated patients. The VNS frequency used in the clinic today is in the range of 10-30 Hz. The evidence for choosing the stimulation frequency is limited, and little knowledge is available on the effect of other VNS frequencies. Deep brain, trigeminal nerve, or spinal cord stimulation studies have suggested the use of stimulation frequencies above 80 Hz for seizure control. Therefore, our objective for the present study was to investigate if VNS using frequencies higher than those currently used in the clinic could be more effective in attenuating seizures. Spike-and-wave (SW) discharges were induced in 11 rats, which then were subjected to VNS sessions applied at the frequencies of 10, 30, 80, 130, and 180 Hz combined with control intervals without stimulation. The anticonvulsive effect of VNS was evaluated by comparing the normalized mean power (nMP) and frequency (nMSF) of the SW discharges derived from intracortical recordings collected during the stimulation and control intervals. Compared with the control intervals, all the tested VNS frequencies significantly reduced the nMP (in the range of 9-21%). However, we found that 130 and 180 Hz VNS induced a 50% larger attenuation of seizures than that achieved by 30 Hz VNS. In addition, we found that 80, 130, and 180 Hz VNS induced a significant reduction of the nMSF, that is by 5, 7, and 8%, respectively. These results suggest that a VNS stimulation frequency in the range of 130-180 Hz may be more effective in inhibiting seizures than the 30 Hz VNS applied in the clinic today.