FiberSecure suture compared to braided polyester suture.
Melvin, Alan J; Litsky, Alan S; Juncosa-Melvin, Natalia.
Afiliación
  • Melvin AJ; Surgical Energetics LLC, Covington, Kentucky.
  • Litsky AS; Ohio State University, Columbus, Ohio.
  • Juncosa-Melvin N; Surgical Energetics LLC, Covington, Kentucky.
J Biomed Mater Res B Appl Biomater ; 105(5): 1126-1130, 2017 07.
Article en En | MEDLINE | ID: mdl-27015658
ABSTRACT
Reliability of wound closure is limited primarily by the capacity of tissues to support conventional sutures (or staples), not by strength of either material per se. We developed FiberSecure™ for closures to surpass tissue strength. We assessed and compared the mechanical and histological performance of FiberSecure™ suture versus commercially available braided polyester suture (Mersilene) in the closure of abdominal muscle incisions in miniature swine at approximately 3 months postsurgery. Four incisions were closed in the external oblique muscle of eight Sinclair minipigs. Two wounds were closed with FiberSecure™ suture size 0 and the remaining two with Mersilene suture size 0. At 90 days, specimens were removed for biomechanics and histology. In destructive tensile testing, in the 16 abdominal muscle specimens for the FiberSecure™ suture, muscle tear was not near the suture implantation region, which remained intact. Wound strength met or exceeded strength of neighboring tissue in FiberSecure™ groups, which had peak force of 55.7 ± 22.1 N (mean ± SD) and peak stress of 579.0 ± 159.2 KPa (mean ± SD). For Mersilene, 3 of the 16 samples tore at the suture site and the remaining samples tore through the abdominal muscle not near the implantation region. The wound strength was similar to surrounding tissue, and these specimens had peak force of 51.8 ± 21.7 N and peak stress of 550.3 ± 239.4 KPa (mean ± SD). No significant difference was observed in peak force or stress between groups (p > 0.05), most repairs having met or exceeded native tissue strength by this time point. © 2016 Wiley Periodicals, Inc. J Biomed Mater Res Part B Appl Biomater, 105B 1126-1130, 2017.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Poliésteres / Suturas / Ensayo de Materiales / Técnicas de Cierre de Heridas Límite: Animals Idioma: En Revista: J Biomed Mater Res B Appl Biomater Asunto de la revista: ENGENHARIA BIOMEDICA Año: 2017 Tipo del documento: Article

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Poliésteres / Suturas / Ensayo de Materiales / Técnicas de Cierre de Heridas Límite: Animals Idioma: En Revista: J Biomed Mater Res B Appl Biomater Asunto de la revista: ENGENHARIA BIOMEDICA Año: 2017 Tipo del documento: Article