Source strength verification and quality assurance of preloaded brachytherapy needles using a CMOS flat panel detector.
Golshan, Maryam; Spadinger, Ingrid; Chng, Nick.
Afiliación
  • Golshan M; Department of Physics, University of British Columbia, Vancouver, British Columbia V6T1Z1, Canada and Department of Medical Physics, Vancouver Center, British Columbia Cancer Agency, Vancouver, British Columbia V5Z 4E6, Canada.
  • Spadinger I; Department of Medical Physics, Vancouver Center, British Columbia Cancer Agency, Vancouver, British Columbia V5Z 4E6, Canada.
  • Chng N; Department of Medical Physics, Center for the North, British Columbia Cancer Agency, Prince George, British Columbia V2M 7E9, Canada.
Med Phys ; 43(6): 3008-3018, 2016 Jun.
Article en En | MEDLINE | ID: mdl-27277049
PURPOSE: Current methods of low dose rate brachytherapy source strength verification for sources preloaded into needles consist of either assaying a small number of seeds from a separate sample belonging to the same lot used to load the needles or performing batch assays of a subset of the preloaded seed trains. Both of these methods are cumbersome and have the limitations inherent to sampling. The purpose of this work was to investigate an alternative approach that uses an image-based, autoradiographic system capable of the rapid and complete assay of all sources without compromising sterility. METHODS: The system consists of a flat panel image detector, an autoclavable needle holder, and software to analyze the detected signals. The needle holder was designed to maintain a fixed vertical spacing between the needles and the image detector, and to collimate the emissions from each seed. It also provides a sterile barrier between the needles and the imager. The image detector has a sufficiently large image capture area to allow several needles to be analyzed simultaneously.Several tests were performed to assess the accuracy and reproducibility of source strengths obtained using this system. Three different seed models (Oncura 6711 and 9011 (125)I seeds, and IsoAid Advantage (103)Pd seeds) were used in the evaluations. Seeds were loaded into trains with at least 1 cm spacing. RESULTS: Using our system, it was possible to obtain linear calibration curves with coverage factor k = 1 prediction intervals of less than ±2% near the centre of their range for the three source models. The uncertainty budget calculated from a combination of type A and type B estimates of potential sources of error was somewhat larger, yielding (k = 1) combined uncertainties for individual seed readings of 6.2% for (125)I 6711 seeds, 4.7% for (125)I 9011 seeds, and 11.0% for Advantage (103)Pd seeds. CONCLUSIONS: This study showed that a flat panel detector dosimetry system is a viable option for source strength verification in preloaded needles, as it is capable of measuring all of the sources intended for implantation. Such a system has the potential to directly and efficiently estimate individual source strengths, the overall mean source strength, and the positions within the seed-spacer train.

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Tipo de estudio: Prognostic_studies Idioma: En Revista: Med Phys Año: 2016 Tipo del documento: Article País de afiliación: Canadá Pais de publicación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Tipo de estudio: Prognostic_studies Idioma: En Revista: Med Phys Año: 2016 Tipo del documento: Article País de afiliación: Canadá Pais de publicación: Estados Unidos