Prediction of hip osteoporosis by DXA using a novel pulse-echo ultrasound device.
Schousboe, J T; Riekkinen, O; Karjalainen, J.
Afiliación
  • Schousboe JT; HealthPartners Institute and Park Nicollet Clinic, 3800 Park Nicollet Blvd, Minneapolis, MN, 55416, USA. schouj@parknicollet.com.
  • Riekkinen O; Division of Health Policy and Management, University of Minnesota, Minneapolis, USA. schouj@parknicollet.com.
  • Karjalainen J; Bone Index Finland, Ltd, Kuopio, Finland.
Osteoporos Int ; 28(1): 85-93, 2017 01.
Article en En | MEDLINE | ID: mdl-27492435
ABSTRACT
Pulse-echo ultrasonometry can be used as a pre-screen for hip osteoporosis before dual-energy x-ray absorptiometry (DXA), potentially allowing DXA to be avoided for the majority of post-menopausal women. Pulse-echo ultrasound measures of tibia cortical thickness are also associated with radiographically confirmed prior fractures, independent of femoral neck bone mineral density.

INTRODUCTION:

To estimate how well a pulse-echo ultrasound device discriminates those who have from those who do not have hip osteoporosis (femoral neck bone mineral density [BMD] or total hip BMD T-score ≤ -2.5), and to estimate the association of pulse-echo ultrasound measures with prevalent (radiographically confirmed) clinical fractures.

METHODS:

Five hundred fifty-five post-menopausal women age 50 to 89 had femoral neck and total hip BMD measured by dual-energy x-ray absorptiometry (DXA), and pulse-echo ultrasound measures of distal radius, proximal tibia, distal tibia cortical thickness, and multi- and single-site density indices (DI). Using previously published threshold ultrasound values, we estimated the proportion of women who would avoid a follow-up DXA after pulse-echo ultrasonometry, and the sensitivity and specificity of this for the detection of hip osteoporosis. Logistic regression models were used to estimate the associations of pulse-echo ultrasound measures with radiographically confirmed clinical fractures within the prior 5 years.

RESULTS:

Using multi-site and single-site DI measures, follow-up DXA could be avoided for 73 and 69 % of individuals, respectively, while detecting hip osteoporosis with 80-82 % sensitivity and 81 % specificity. Radiographically confirmed prior fracture was associated with ultrasound measures of single-site DI (odds ratio (OR) 1.55, 95 % confidence interval (CI). 1.06 to 2.26) and proximal tibia cortical thickness (OR 1.47, 95 % CI 1.10 to 1.96), adjusted for age, body mass index, and femoral neck BMD.

CONCLUSIONS:

Pulse-echo ultrasonometry can be used as an initial screening test for hip osteoporosis. Prospective studies of how well pulse-echo ultrasound measures predict subsequent clinical fractures are warranted.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Osteoporosis Posmenopáusica / Articulación de la Cadera Tipo de estudio: Diagnostic_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Límite: Aged / Aged80 / Female / Humans / Middle aged Idioma: En Revista: Osteoporos Int Asunto de la revista: METABOLISMO / ORTOPEDIA Año: 2017 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Osteoporosis Posmenopáusica / Articulación de la Cadera Tipo de estudio: Diagnostic_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Límite: Aged / Aged80 / Female / Humans / Middle aged Idioma: En Revista: Osteoporos Int Asunto de la revista: METABOLISMO / ORTOPEDIA Año: 2017 Tipo del documento: Article País de afiliación: Estados Unidos
...