Your browser doesn't support javascript.
loading
Novel use of the medtronic micro vascular plug for PDA closure in preterm infants.
Wang-Giuffre, Elizabeth W; Breinholt, John P.
Afiliación
  • Wang-Giuffre EW; Division of Pediatric Cardiology, Department of Pediatrics, University of Texas Health Science Center at Houston, Houston, Texas.
  • Breinholt JP; Division of Pediatric Cardiology, Department of Pediatrics, University of Texas Health Science Center at Houston, Houston, Texas.
Catheter Cardiovasc Interv ; 89(6): 1059-1065, 2017 May.
Article en En | MEDLINE | ID: mdl-27860173
ABSTRACT

OBJECTIVES:

To describe a single institution experience with a new endovascular occlusion device used for transcatheter patent ductus arteriosus (PDA) closure in preterm infants.

BACKGROUND:

The PDA is a defect largely treated via cardiac catheterization except for the smallest patients.

METHODS:

Medical records and catheterization reports of all premature neonates who underwent PDA device closure with the Medtronic Micro Vascular Plug (MVP) (Medtronic, Minneapolis, MN) between September 2015 and June 2016 were reviewed. Procedural details, complications, and short term outcomes were recorded.

RESULTS:

Eight premature neonates born at a median gestational age of 28 weeks (23 to 35 weeks) underwent PDA closure with a Medtronic MVP. All devices were deployed via a 4F angled Glide catheter in prograde fashion without arterial access. Median age and weight was 52 days (15-112 days) and 2,550 g (1,800-3,500 g), respectively. Fluoroscopy and echocardiography were utilized for the procedure. Complete closure was achieved in all patients with no procedural complications, pulmonary artery or aortic obstruction or death. One device embolized 9 days after deployment and was successfully retrieved.

CONCLUSIONS:

This study describes transcatheter PDA closure with the Medtronic MVP. The major advantage is the ability to deliver the device via a standard catheter in prograde fashion, and avoid the hemodynamic instability produced by rigid delivery sheaths. It also further validates the ability to close the PDA without arterial access, providing improved access to this procedure to smaller and more vulnerable children. © 2016 Wiley Periodicals, Inc.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Cateterismo Cardíaco / Conducto Arterioso Permeable / Dispositivo Oclusor Septal / Recien Nacido Extremadamente Prematuro Límite: Female / Humans / Male / Newborn País/Región como asunto: America do norte Idioma: En Revista: Catheter Cardiovasc Interv Asunto de la revista: CARDIOLOGIA Año: 2017 Tipo del documento: Article

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Cateterismo Cardíaco / Conducto Arterioso Permeable / Dispositivo Oclusor Septal / Recien Nacido Extremadamente Prematuro Límite: Female / Humans / Male / Newborn País/Región como asunto: America do norte Idioma: En Revista: Catheter Cardiovasc Interv Asunto de la revista: CARDIOLOGIA Año: 2017 Tipo del documento: Article
...