Autoclave Sterilization of PEDOT:PSS Electrophysiology Devices.
Adv Healthc Mater
; 5(24): 3094-3098, 2016 12.
Article
en En
| MEDLINE
| ID: mdl-27885829
ABSTRACT
Autoclaving, the most widely available sterilization method, is applied to poly(3,4-ethylenedioxythiophene) doped with polystyrene sulfonate (PEDOTPSS) electrophysiology devices. The process does not harm morphology or electrical properties, while it effectively kills E. coli intentionally cultured on the devices. This finding paves the way to widespread introduction of PEDOTPSS electrophysiology devices to the clinic.