Your browser doesn't support javascript.
loading
Novel centrifugal technology for measuring sperm concentration in the home.
Schaff, Ulrich Y; Fredriksen, Laura L; Epperson, Jon G; Quebral, Tiffany R; Naab, Sara; Sarno, Mark J; Eisenberg, Michael L; Sommer, Greg J.
Afiliación
  • Schaff UY; Sandstone Diagnostics, Inc., Livermore, California.
  • Fredriksen LL; Sandstone Diagnostics, Inc., Livermore, California.
  • Epperson JG; Sandstone Diagnostics, Inc., Livermore, California.
  • Quebral TR; Sandstone Diagnostics, Inc., Livermore, California.
  • Naab S; Sandstone Diagnostics, Inc., Livermore, California.
  • Sarno MJ; Vision Biotechnology Consulting, Escondido, California.
  • Eisenberg ML; Department of Urology and Department of Obstetrics & Gynecology, Stanford University School of Medicine, Palo Alto, California.
  • Sommer GJ; Sandstone Diagnostics, Inc., Livermore, California. Electronic address: gsommer@sandstonedx.com.
Fertil Steril ; 107(2): 358-364.e4, 2017 02.
Article en En | MEDLINE | ID: mdl-27887718
ABSTRACT

OBJECTIVE:

To evaluate the analytical performance and usability of the Trak Male Fertility Testing System, a semiquantitative (categorical) device recently US Food and Drug Administration (FDA)-cleared for measuring sperm concentration in the home by untrained users.

DESIGN:

A three-site clinical trial comparing self-reported lay user results versus reference results obtained by computer-aided semen analysis (CASA).

SETTING:

Simulated home use environments at fertility centers and urologist offices. PATIENT(S) A total of 239 untrained users. INTERVENTION(S) None. MAIN OUTCOME MEASURE(S) Sperm concentration results reported from self-testing lay users and laboratory reference method by CASA were evaluated semiquantitatively against the device's clinical cutoffs of 15 M/mL (current World Health Organization cutoff) and 55 M/mL (associated with faster time to pregnancy). Additional reported metrics include assay linearity, precision, limit of detection, and ease-of-use ratings from lay users. RESULT(S) Lay users achieved an accuracy (versus the reference) of 93.3% (95% confidence interval [CI] 84.1%-97.4%) for results categorized as ≤15 M/mL, 82.4% (95% CI 73.3%-88.9%) for results categorized as 15-55 M/mL, and 95.5% (95% CI 88.9%-98.2%) for results categorized as >55 M/mL. When measured quantitatively, Trak results had a strong linear correlation with CASA measurements (r = 0.99). The precision and limit of detection studies show that the device has adequate reproducibility and detection range for home use. Subjects generally rated the device as easy to use. CONCLUSION(S) The Trak System is an accurate tool for semiquantitatively measuring sperm concentration in the home. The system may enable screening and longitudinal assessment of sperm concentration at home. CLINICAL TRIAL REGISTRATION NUMBER ClinicalTrials.gov identifier NCT02475395.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Autocuidado / Recuento de Espermatozoides / Espermatozoides / Centrifugación / Fertilidad / Infertilidad Masculina Tipo de estudio: Clinical_trials / Etiology_studies / Prognostic_studies Límite: Adult / Humans / Male / Middle aged País/Región como asunto: America do norte Idioma: En Revista: Fertil Steril Año: 2017 Tipo del documento: Article

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Autocuidado / Recuento de Espermatozoides / Espermatozoides / Centrifugación / Fertilidad / Infertilidad Masculina Tipo de estudio: Clinical_trials / Etiology_studies / Prognostic_studies Límite: Adult / Humans / Male / Middle aged País/Región como asunto: America do norte Idioma: En Revista: Fertil Steril Año: 2017 Tipo del documento: Article