OBJECTIVE:

To investigate clinical outcomes and subjective ratings of the Kanso™ off-the-ear (OTE) cochlear implant sound processor.

DESIGN:

Prospective, within-subject design investigating outcomes with a range of single and dual-microphone programmes for Kanso compared to conventional behind-the-ear (BTE) sound processors. STUDY SAMPLE Twenty post-lingually hearing-impaired cochlear implant recipients who were experienced Nucleus® 5 or Nucleus® 6 BTE users.

RESULTS:

No significant difference in performance was found for words in quiet or sentences in co-located noise between the Kanso and Nucleus 6 devices. For the moderately directional Standard programme, no significant difference was found for sentences in spatially separated noise between the Kanso and Nucleus 6 devices, but a performance decrement between 1.4 and 2.0 dB was found in highly directional and adaptive directional programmes. The default Kanso programme, SCAN, provided improvements of 6.9 dB over a single-microphone programme and 2.3 dB over the Standard programme in spatially separated noise. Participants rated Kanso significantly better than their own BTE processor on measures of comfort, look and feel, ease of use, music and overall hearing performance.

CONCLUSION:

Dual-microphone directional processing provides significant benefit over a single microphone for OTE processors. This study demonstrates clinical outcomes and acceptance of the Kanso OTE sound processor.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Percepción del Habla / Implantes Cocleares / Personas con Deficiencia Auditiva / Implantación Coclear / Pérdida Auditiva Tipo de estudio: Diagnostic_studies / Observational_studies Límite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Int J Audiol Asunto de la revista: AUDIOLOGIA Año: 2017 Tipo del documento: Article País de afiliación: Australia

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Percepción del Habla / Implantes Cocleares / Personas con Deficiencia Auditiva / Implantación Coclear / Pérdida Auditiva Tipo de estudio: Diagnostic_studies / Observational_studies Límite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Int J Audiol Asunto de la revista: AUDIOLOGIA Año: 2017 Tipo del documento: Article País de afiliación: Australia