Is it possible to avoid intra-articular screw penetration with minimal use of fluoroscopy in the application of distal radius volar plate?
Aldemir, Cengiz; Duygun, Fatih.
Afiliación
  • Aldemir C; Department of Orthopedics and Traumatology, Saglik Bilimleri University, Antalya Training and Research Hospital, 07030 Muratpasa, Antalya, Turkey. aldemircengiz@yahoo.com.
Eklem Hastalik Cerrahisi ; 28(1): 2-6, 2017 Apr.
Article en En | MEDLINE | ID: mdl-28291431
ABSTRACT

OBJECTIVES:

This study aims to investigate whether or not it is possible to avoid intra-articular screw penetration in the application of distal radius volar plate with minimal use of fluoroscopy and thereby reduce to a minimum the risk of exposure of the surgical team to radiation. PATIENTS AND

METHODS:

Volar plate with distal screw application with minimal use of fluoroscopy was performed in a total of 50 patients (22 males, 28 females; mean age 43.52 years; range 18 to 76 years) with unstable distal radius fracture following reduction of fracture between June 2014 and May 2016. A specially designed volar anatomic plate was applied by creating a 3 mm gap proximal from the tip of the facet of the lunate fossa.

RESULTS:

No intra-articular screw complication was observed during plate-screw applications that we performed by using fluoroscopy at a minimum level and paying attention to the surgical technique described with the plate designed by giving a specific angle to screw beds. During the operation, fluoroscopy was performed at a minimum level and no finding was observed which could suggest use of intra-articular screw. No findings of restriction in wrist movements, locking or friction were observed intra- or postoperatively.

CONCLUSION:

With improvements in the design of the volar plate and the distal locking screw bed, it is possible to rule out the possibility of intra-articular screw complications and the need for surgical re-correction in the early-term, and to reduce to a minimum the risk of exposure of the surgical team to radiation.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Ortopedia / Fracturas del Radio / Tornillos Óseos / Exposición Profesional / Exposición a la Radiación / Fijación Interna de Fracturas Límite: Adolescent / Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Eklem Hastalik Cerrahisi Asunto de la revista: ORTOPEDIA Año: 2017 Tipo del documento: Article País de afiliación: Turquía

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Ortopedia / Fracturas del Radio / Tornillos Óseos / Exposición Profesional / Exposición a la Radiación / Fijación Interna de Fracturas Límite: Adolescent / Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Eklem Hastalik Cerrahisi Asunto de la revista: ORTOPEDIA Año: 2017 Tipo del documento: Article País de afiliación: Turquía