Evaluating the accuracy of 4D-CT ventilation imaging: First comparison with Technegas SPECT ventilation.
Hegi-Johnson, Fiona; Keall, Paul; Barber, Jeff; Bui, Chuong; Kipritidis, John.
Afiliación
  • Hegi-Johnson F; Radiation Physics Laboratory, Faculty of Medicine, Sydney University, Camperdown, NSW, 2006, Australia.
  • Keall P; Department of Medical Physics, School of Mathematical and Physical Sciences, University of Newcastle, Newcastle, NSW, 2300, Australia.
  • Barber J; Radiation Oncology Centre, Seventh Day Adventist Hospital, Wahroonga, NSW 2076, Australia.
  • Bui C; Department of Radiation Oncology, Sir Peter MacCallum Department of Oncology, University of Melbourne, Melbourne, Vic., 3000, Australia.
  • Kipritidis J; Radiation Physics Laboratory, Faculty of Medicine, Sydney University, Camperdown, NSW, 2006, Australia.
Med Phys ; 44(8): 4045-4055, 2017 Aug.
Article en En | MEDLINE | ID: mdl-28477378
ABSTRACT

PURPOSE:

Computed tomography ventilation imaging (CTVI) is a highly accessible functional lung imaging modality that can unlock the potential for functional avoidance in lung cancer radiation therapy. Previous attempts to validate CTVI against clinical ventilation single-photon emission computed tomography (V-SPECT) have been hindered by radioaerosol clumping artifacts. This work builds on those studies by performing the first comparison of CTVI with 99m Tc-carbon ('Technegas'), a clinical V-SPECT modality featuring smaller radioaerosol particles with less clumping.

METHODS:

Eleven lung cancer radiotherapy patients with early stage (T1/T2N0) disease received treatment planning four-dimensional CT (4DCT) scans paired with Technegas V/Q-SPECT/CT. For each patient, we applied three different CTVI methods. Two of these used deformable image registration (DIR) to quantify breathing-induced lung density changes (CTVIDIR-HU ), or breathing-induced lung volume changes (CTVIDIR-Jac ) between the 4DCT exhale/inhale phases. A third method calculated the regional product of air-tissue densities (CTVIHU ) and did not involve DIR. Corresponding CTVI and V-SPECT scans were compared using the Dice similarity coefficient (DSC) for functional defect and nondefect regions, as well as the Spearman's correlation r computed over the whole lung. The DIR target registration error (TRE) was quantified using both manual and computer-selected anatomic landmarks.

RESULTS:

Interestingly, the overall best performing method (CTVIHU ) did not involve DIR. For nondefect regions, the CTVIHU , CTVIDIR-HU , and CTVIDIR-Jac methods achieved mean DSC values of 0.69, 0.68, and 0.54, respectively. For defect regions, the respective DSC values were moderate 0.39, 0.33, and 0.44. The Spearman r-values were generally weak 0.26 for CTVIHU , 0.18 for CTVIDIR-HU , and -0.02 for CTVIDIR-Jac . The spatial accuracy of CTVI was not significantly correlated with TRE, however the DIR accuracy itself was poor with TRE > 3.6 mm on average, potentially indicative of poor quality 4DCT. Q-SPECT scans achieved good correlations with V-SPECT (mean r > 0.6), suggesting that the image quality of Technegas V-SPECT was not a limiting factor in this study.

CONCLUSIONS:

We performed a validation of CTVI using clinically available 4DCT and Technegas V/Q-SPECT for 11 lung cancer patients. The results reinforce earlier findings that the spatial accuracy of CTVI exhibits significant interpatient and intermethod variability. We propose that the most likely factor affecting CTVI accuracy was poor image quality of clinical 4DCT.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Tomografía Computarizada de Emisión de Fotón Único / Pertecnetato de Sodio Tc 99m / Ventilación Pulmonar / Tomografía Computarizada Cuatridimensional Tipo de estudio: Guideline Límite: Humans Idioma: En Revista: Med Phys Año: 2017 Tipo del documento: Article País de afiliación: Australia Pais de publicación: EEUU / ESTADOS UNIDOS / ESTADOS UNIDOS DA AMERICA / EUA / UNITED STATES / UNITED STATES OF AMERICA / US / USA

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Tomografía Computarizada de Emisión de Fotón Único / Pertecnetato de Sodio Tc 99m / Ventilación Pulmonar / Tomografía Computarizada Cuatridimensional Tipo de estudio: Guideline Límite: Humans Idioma: En Revista: Med Phys Año: 2017 Tipo del documento: Article País de afiliación: Australia Pais de publicación: EEUU / ESTADOS UNIDOS / ESTADOS UNIDOS DA AMERICA / EUA / UNITED STATES / UNITED STATES OF AMERICA / US / USA