Your browser doesn't support javascript.
loading
Poster - Thur Eve - 68: Evaluation and analytical comparison of different 2D and 3D treatment planning systems using dosimetry in anthropomorphic phantom.
Khosravi, H R; Nodehi, Mr Golrokh; Asnaashari, Kh; Mahdavi, S R; Shirazi, A R; Gholami, S.
Afiliación
  • Khosravi HR; Nuclear Sciences and Technology Research Institute.
  • Nodehi MG; Science and Research branch, Islamic Azad University.
  • Asnaashari K; Tehran University of Medical Sciences.
  • Mahdavi SR; Cancer Institute of Imam Khomeini hospital.
  • Shirazi AR; Tehran University of Medical Sciences.
  • Gholami S; Tehran University of Medical Sciences.
Med Phys ; 39(7Part4): 4637-4638, 2012 Jul.
Article en En | MEDLINE | ID: mdl-28516655
ABSTRACT

PURPOSE:

The aim of this study was to evaluate and analytically compare different calculation algorithms applied in our country radiotherapy centers base on the methodology developed by IAEA for treatment planning systems (TPS) commissioning (IAEA TEC-DOC 1583). MATERIAL &

METHODS:

Thorax anthropomorphic phantom (002LFC CIRS inc.), was used to measure 7 tests that simulate the whole chain of external beam TPS. The dose were measured with ion chambers and the deviation between measured and TPS calculated dose was reported. This methodology, which employs the same phantom and the same setup test cases, was tested in 4 different hospitals which were using 5 different algorithms/ inhomogeneity correction methods implemented in different TPS. The algorithms in this study were divided into two groups including correction based and model based algorithms.

RESULTS:

A total of 84 clinical test case datasets for different energies and calculation algorithms were produced, which amounts of differences in inhomogeneity points with low density (lung) and high density (bone) was decreased meaningfully with advanced algorithms. The number of deviations outside agreement criteria was increased with the beam energy and decreased with advancement of the TPS calculation algorithm.

CONCLUSION:

Large deviations were seen in some correction based algorithms, so sophisticated algorithms, would be preferred in clinical practices, especially for calculation in inhomogeneous media. Use of model based algorithms with lateral transport calculation, is recommended. Some systematic errors which were revealed during this study, is showing necessity of performing periodic audits on TPS in radiotherapy centers.
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Tipo de estudio: Prognostic_studies Idioma: En Revista: Med Phys Año: 2012 Tipo del documento: Article Pais de publicación: EEUU / ESTADOS UNIDOS / ESTADOS UNIDOS DA AMERICA / EUA / UNITED STATES / UNITED STATES OF AMERICA / US / USA

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Tipo de estudio: Prognostic_studies Idioma: En Revista: Med Phys Año: 2012 Tipo del documento: Article Pais de publicación: EEUU / ESTADOS UNIDOS / ESTADOS UNIDOS DA AMERICA / EUA / UNITED STATES / UNITED STATES OF AMERICA / US / USA