Diagnostic accuracy of contrast-enhanced ultrasound (CEUS) in monitoring vascular complications in patients after liver transplantation - diagnostic performance compared with histopathological results.
Rübenthaler, J; Paprottka, K J; Hameister, E; Hoffmann, K; Joiko, N; Reiser, M; Rjosk-Dendorfer, D; Clevert, D A.
Afiliación
  • Rübenthaler J; Department of Clinical Radiology, Ludwig-Maximilians-University of Munich-Grosshadern Campus, Munich, Germany.
  • Paprottka KJ; Department of Clinical Radiology, Ludwig-Maximilians-University of Munich-Grosshadern Campus, Munich, Germany.
  • Hameister E; Department of Clinical Radiology, Ludwig-Maximilians-University of Munich-Grosshadern Campus, Munich, Germany.
  • Hoffmann K; Department of Clinical Radiology, Ludwig-Maximilians-University of Munich-Grosshadern Campus, Munich, Germany.
  • Joiko N; Department of Clinical Radiology, Ludwig-Maximilians-University of Munich-Grosshadern Campus, Munich, Germany.
  • Reiser M; Department of Clinical Radiology, Ludwig-Maximilians-University of Munich-Grosshadern Campus, Munich, Germany.
  • Rjosk-Dendorfer D; Department of Clinical Radiology, Ludwig-Maximilians-University of Munich-Grosshadern Campus, Munich, Germany.
  • Clevert DA; Department of Clinical Radiology, Ludwig-Maximilians-University of Munich-Grosshadern Campus, Munich, Germany.
Clin Hemorheol Microcirc ; 66(4): 311-316, 2017.
Article en En | MEDLINE | ID: mdl-28527202
ABSTRACT

PURPOSE:

To analyse the diagnostic performance of contrast-enhanced ultrasound (CEUS) in patients with vascular complications and transplant rejection compared to histopathological results. MATERIALS AND

METHODS:

Our study consisted of 45 retrospectively analysed patients that underwent liver transplantations between January 1993 and December 2015 and developed post-transplant vascular complications with transplant rejection. CEUS examinations took place between September 2006 and December 2015. CEUS findings were correlated with histopathological results.

RESULTS:

CEUS showed a sensitivity of 61.5%, a specificity of 100.0%, a positive predictive value (PPV) of 100.0% and a negative predictive value (NPV) of 86,5% in the detection of vascular complications with post-transplant rejection. 5 examinations were reported as normal whereas the histopathological result showed a transplant rejection (false-negative).

CONCLUSION:

CEUS might be a useful additional non-invasive technique for the assessment of vascular complications with post-transplant rejection in patients after liver transplantation.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Ultrasonografía / Medios de Contraste Tipo de estudio: Diagnostic_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Límite: Adolescent / Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Clin Hemorheol Microcirc Asunto de la revista: ANGIOLOGIA / HEMATOLOGIA Año: 2017 Tipo del documento: Article País de afiliación: Alemania

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Ultrasonografía / Medios de Contraste Tipo de estudio: Diagnostic_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Límite: Adolescent / Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Clin Hemorheol Microcirc Asunto de la revista: ANGIOLOGIA / HEMATOLOGIA Año: 2017 Tipo del documento: Article País de afiliación: Alemania