Performance Evaluation of Serum Free Light Chain Analysis: Nephelometry vs Turbidimetry, Monoclonal vs Polyclonal Reagents.
Messiaen, Anne-Sophie; De Sloovere, Maxime M W; Claus, Paul-Emile; Vercammen, Martine; Van Hoovels, Lieve; Heylen, Olivier; Debrabandere, Johan; Vanpoucke, Hilde; De Smet, Dieter.
Afiliación
  • Messiaen AS; From the Department of Laboratory Medicine, AZ Delta Roeselare-Menen, Roeselare, Belgium.
  • De Sloovere MMW; From the Department of Laboratory Medicine, AZ Delta Roeselare-Menen, Roeselare, Belgium.
  • Claus PE; From the Department of Laboratory Medicine, AZ Delta Roeselare-Menen, Roeselare, Belgium.
  • Vercammen M; Department of Laboratory Medicine, AZ Sint-Jan Brugge-Oostende, Brugge, Belgium.
  • Van Hoovels L; Department of Laboratory Medicine, OLV Aalst-Asse, Aalst, Belgium; and.
  • Heylen O; Department of Laboratory Medicine, AZ Groeninge, Kortrijk, Belgium.
  • Debrabandere J; From the Department of Laboratory Medicine, AZ Delta Roeselare-Menen, Roeselare, Belgium.
  • Vanpoucke H; From the Department of Laboratory Medicine, AZ Delta Roeselare-Menen, Roeselare, Belgium.
  • De Smet D; From the Department of Laboratory Medicine, AZ Delta Roeselare-Menen, Roeselare, Belgium.
Am J Clin Pathol ; 147(6): 611-622, 2017 Jun 01.
Article en En | MEDLINE | ID: mdl-28575180
OBJECTIVES: Free light chain (FLC) measurement gained a lot of interest for diagnostic workup of monoclonal gammopathy. METHODS: We evaluated the performance of turbidimetric polyclonal Freelite (The Binding Site, Birmingham, UK) assays on Cobas 6000 (Roche Diagnostics, Rotkreuz, Switzerland) and nephelometric monoclonal N Latex (Siemens Healthcare Diagnostics, Marburg, Germany) assays on BN ProSpec (Dade Behring, Deerfield, IL) vs established nephelometric Freelite assays on BN ProSpec. RESULTS: Analytical performance was acceptable. Method comparison (n = 118) showed significant proportional FLC differences for N Latex assays. However, good correlation and clinical concordance were shown. Recovery study in the low concentration range demonstrated consistent over- and underrecovery for Freelite reagents, hampering future research on prognostic value of suppressed noninvolved FLC. Antigen excess detection was successful for κ FLC in three-fourths of cases with Freelite reagents and in all cases with N Latex reagents. However, the latter resulted in underestimated κ FLC concentrations. CONCLUSIONS: FLC analysis requires continuous awareness of analytical limitations. Monitoring of disease response requires FLC analysis on the same platform using the same reagents.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Paraproteinemias / Cadenas kappa de Inmunoglobulina / Cadenas lambda de Inmunoglobulina / Cadenas Ligeras de Inmunoglobulina / Nefelometría y Turbidimetría Tipo de estudio: Diagnostic_studies / Evaluation_studies / Prognostic_studies Límite: Humans Idioma: En Revista: Am J Clin Pathol Año: 2017 Tipo del documento: Article País de afiliación: Bélgica Pais de publicación: Reino Unido

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Paraproteinemias / Cadenas kappa de Inmunoglobulina / Cadenas lambda de Inmunoglobulina / Cadenas Ligeras de Inmunoglobulina / Nefelometría y Turbidimetría Tipo de estudio: Diagnostic_studies / Evaluation_studies / Prognostic_studies Límite: Humans Idioma: En Revista: Am J Clin Pathol Año: 2017 Tipo del documento: Article País de afiliación: Bélgica Pais de publicación: Reino Unido