Your browser doesn't support javascript.
loading
Multicenter comparison of the new Cobas 6800 system with Cobas Ampliprep/Cobas TaqMan and Abbott RealTime for the quantification of HIV, HBV and HCV viral load.
Wirden, Marc; Larrouy, Lucile; Mahjoub, Nadia; Todesco, Eve; Damond, Florence; Delagreverie, Heloise; Akhavan, Sepideh; Charpentier, Charlotte; Chaix, Marie-Laure; Descamps, Diane; Calvez, Vincent; Marcelin, Anne-Genevieve.
Afiliación
  • Wirden M; Sorbonne Universités, UPMC Univ Paris 06, INSERM, Institut Pierre Louis d'épidémiologie et de Santé Publique (IPLESP UMRS 1136), F75013 Paris, France; AP-HP, Groupe Hospitalier Pitié-Salpêtrière, Service de Virologie, Paris, France. Electronic address: marc.wirden@aphp.fr.
  • Larrouy L; IAME, UMR1137, INSERM, Université Paris Diderot Sorbonne Paris Cité, AP-HP, Laboratoire de Virologie, Hôpital Bichat-Claude Bernard, Paris, France.
  • Mahjoub N; INSERM U941, Université Paris Diderot, Laboratoire de Virologie, AP-HP, Hôpital Saint-Louis, Paris, France.
  • Todesco E; Sorbonne Universités, UPMC Univ Paris 06, INSERM, Institut Pierre Louis d'épidémiologie et de Santé Publique (IPLESP UMRS 1136), F75013 Paris, France; AP-HP, Groupe Hospitalier Pitié-Salpêtrière, Service de Virologie, Paris, France.
  • Damond F; IAME, UMR1137, INSERM, Université Paris Diderot Sorbonne Paris Cité, AP-HP, Laboratoire de Virologie, Hôpital Bichat-Claude Bernard, Paris, France.
  • Delagreverie H; INSERM U941, Université Paris Diderot, Laboratoire de Virologie, AP-HP, Hôpital Saint-Louis, Paris, France.
  • Akhavan S; AP-HP, Groupe Hospitalier Pitié-Salpêtrière, Service de Virologie, Paris, France.
  • Charpentier C; IAME, UMR1137, INSERM, Université Paris Diderot Sorbonne Paris Cité, AP-HP, Laboratoire de Virologie, Hôpital Bichat-Claude Bernard, Paris, France.
  • Chaix ML; INSERM U941, Université Paris Diderot, Laboratoire de Virologie, AP-HP, Hôpital Saint-Louis, Paris, France; CNR VIH, AP-HP, Paris, France.
  • Descamps D; IAME, UMR1137, INSERM, Université Paris Diderot Sorbonne Paris Cité, AP-HP, Laboratoire de Virologie, Hôpital Bichat-Claude Bernard, Paris, France; CNR VIH, AP-HP, Paris, France.
  • Calvez V; Sorbonne Universités, UPMC Univ Paris 06, INSERM, Institut Pierre Louis d'épidémiologie et de Santé Publique (IPLESP UMRS 1136), F75013 Paris, France; AP-HP, Groupe Hospitalier Pitié-Salpêtrière, Service de Virologie, Paris, France.
  • Marcelin AG; Sorbonne Universités, UPMC Univ Paris 06, INSERM, Institut Pierre Louis d'épidémiologie et de Santé Publique (IPLESP UMRS 1136), F75013 Paris, France; AP-HP, Groupe Hospitalier Pitié-Salpêtrière, Service de Virologie, Paris, France.
J Clin Virol ; 96: 49-53, 2017 11.
Article en En | MEDLINE | ID: mdl-28950186
ABSTRACT

BACKGROUND:

The new Roche Cobas 6800 platform (C6800) has been recently introduced for viral load (VL) measurement.

OBJECTIVES:

Comparing C6800 to Cobas Ampliprep/Cobas TaqMan v2.0 (CAP/CTM) for the quantification of HIV, HBV and HCV viremia, and to the Abbott RealTime assay (ABB) for HCV quantification. STUDY

DESIGN:

We analysed 121 samples for HBV, and 139 for HIV-1 including 2 groupO and 137 group M viruses (36.5% subtype B, 27.0% CRF02_AG, 22.6% from other clades, and 14% subtype not available). For the 100 HCV samples compared with CAP/CTM, 42% were genotype 1 and 17% were genotype 4. For the 68 HCV samples compared with ABB, 52.9% were genotype 1 and 22.1% were genotype 4.

RESULTS:

C6800 results correlated well with those of CAP/CTM for all three viruses (R2 0.97-0.99). However, C6800 can yield different viraemia

results:

higher for HIV (mean difference +0.11 log10copies/mL, p<0.0001), and lower for HBV (mean difference-0.11 log10 IU/mL, p<0.0001). Differences exceeded 0.5 log10 for 6.5% of HIV-1 samples and 7.4% of HBV samples. For HCV quantification, C6800 gave mostly lower values than the other assays towards the bottom of the range, and higher values in the upper part of the range, especially in comparisons with ABB, for which 28% of differences exceeded 0.5 log10 IU/mL. No particular HCV genotype was identified as responsible for these differences.

CONCLUSION:

Overall, the comparison tests between C6800 and CAP/CTM systems are satisfactory for the three viruses. Frequent discrepancies were observed between C6800 and ABB for HCV.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Viremia / Virus de la Hepatitis B / VIH / Hepacivirus / Carga Viral / Técnicas de Diagnóstico Molecular Tipo de estudio: Clinical_trials / Diagnostic_studies / Evaluation_studies / Prognostic_studies Límite: Humans Idioma: En Revista: J Clin Virol Asunto de la revista: VIROLOGIA Año: 2017 Tipo del documento: Article Pais de publicación: HOLANDA / HOLLAND / NETHERLANDS / NL / PAISES BAJOS / THE NETHERLANDS

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Viremia / Virus de la Hepatitis B / VIH / Hepacivirus / Carga Viral / Técnicas de Diagnóstico Molecular Tipo de estudio: Clinical_trials / Diagnostic_studies / Evaluation_studies / Prognostic_studies Límite: Humans Idioma: En Revista: J Clin Virol Asunto de la revista: VIROLOGIA Año: 2017 Tipo del documento: Article Pais de publicación: HOLANDA / HOLLAND / NETHERLANDS / NL / PAISES BAJOS / THE NETHERLANDS