Berber, Reshid; Abdel-Gadir, Amna; Rosmini, Stefania; Captur, Gabriella; Nordin, Sabrina; Culotta, Veronica; Palla, Luigi; Kellman, Peter; Lloyd, Guy W; Skinner, John A; Moon, James C; Manisty, Charlotte; Hart, Alister J.
Afiliación
  • Berber R; 1Royal National Orthopaedic Hospital, Institute of Orthopaedics and Musculoskeletal Sciences (R.B., J.A.S., and A.J.H.) and Institute of Cardiovascular Science (A.A.-G., S.N., G.W.L., J.C.M., and C.M.), University College London, London, United Kingdom 2Barts Heart Center, The Cardiovascular Magnetic Resonance Imaging Unit, St Bartholomew's Hospital, London, United Kingdom 3NIHR University College London Hospitals Biomedical Research Centre, London, United Kingdom 4Faculty of Epidemiology and Po
J Bone Joint Surg Am ; 99(21): 1827-1835, 2017 Nov 01.
Article en En | MEDLINE | ID: mdl-29088037
ABSTRACT

BACKGROUND:

High failure rates of metal-on-metal (MoM) hip implants prompted regulatory authorities to issue worldwide safety alerts. Circulating cobalt from these implants causes rare but fatal autopsy-diagnosed cardiotoxicity. There is concern that milder cardiotoxicity may be common and underrecognized. Although blood metal ion levels are easily measured and can be used to track local toxicity, there are no noninvasive tests for organ deposition. We sought to detect correlation between blood metal ions and a comprehensive panel of established markers of early cardiotoxicity.

METHODS:

Ninety patients were recruited into this prospective single-center blinded study. Patients were divided into 3 age and sex-matched groups according to implant type and whole-blood metal ion levels. Group-A patients had a ceramic-on-ceramic [CoC] bearing; Group B, an MoM bearing and low blood metal ion levels; and Group C, an MoM bearing and high blood metal-ion levels. All patients underwent detailed cardiovascular phenotyping using cardiac magnetic resonance imaging (CMR) with T2*, T1, and extracellular volume mapping; echocardiography; and cardiac blood biomarker sampling. T2* is a novel CMR biomarker of tissue metal loading.

RESULTS:

Blood cobalt levels differed significantly among groups A, B, and C (mean and standard deviation [SD], 0.17 ± 0.08, 2.47 ± 1.81, and 30.0 ± 29.1 ppb, respectively) and between group A and groups B and C combined. No significant between-group differences were found in the left atrial or ventricle size, ejection fraction (on CMR or echocardiography), T1 or T2* values, extracellular volume, B-type natriuretic peptide level, or troponin level, and all values were within normal ranges. There was no relationship between cobalt levels and ejection fraction (R = 0.022, 95% confidence interval [CI] = -0.185 to 0.229) or T2* values (R = 0.108, 95% CI = -0.105 to 0.312).

CONCLUSIONS:

Using the best available technologies, we did not find that high (but not extreme) blood cobalt and chromium levels had any significant cardiotoxic effect on patients with an MoM hip implant. There were negligible-to-weak correlations between elevated blood metal ion levels and ejection fraction even at the extremes of the 95% CI, which excludes any clinically important association. LEVEL OF EVIDENCE Therapeutic Level II. See Instructions for Authors for a complete description of levels of evidence.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Imagen por Resonancia Magnética / Artroplastia de Reemplazo de Cadera / Técnicas de Imagen Cardíaca / Prótesis Articulares de Metal sobre Metal / Cardiotoxicidad Tipo de estudio: Etiology_studies / Observational_studies / Prevalence_studies / Risk_factors_studies Límite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: J Bone Joint Surg Am Año: 2017 Tipo del documento: Article

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Imagen por Resonancia Magnética / Artroplastia de Reemplazo de Cadera / Técnicas de Imagen Cardíaca / Prótesis Articulares de Metal sobre Metal / Cardiotoxicidad Tipo de estudio: Etiology_studies / Observational_studies / Prevalence_studies / Risk_factors_studies Límite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: J Bone Joint Surg Am Año: 2017 Tipo del documento: Article