The Patient Size Setting: A Novel Dose Reduction Strategy in Cerebral Endovascular Neurosurgery Using Biplane Fluoroscopy.
Corliss, Brian M; Bennett, Jeffrey; Brennan, Meghan M; Rosemaryam, Azar; Hartman, Cory; Stetler, William R; Polifka, Adam J; Hoh, Brian L; Arreola, Manuel M; Fox, W Christopher.
Afiliación
  • Corliss BM; Department of Neurological Surgery, University of Florida, Gainesville, Florida, USA. Electronic address: Brianmatthew.corliss@neurosurgery.ufl.edu.
  • Bennett J; Department of Radiology, University of Arizona, Tucson, Arizona, USA.
  • Brennan MM; Department of Anesthesiology, University of Florida, Gainesville, Florida, USA.
  • Rosemaryam A; Department of Radiology, University of Florida, Gainesville, Florida, USA.
  • Hartman C; Department of Neurological Surgery, University of Florida, Gainesville, Florida, USA.
  • Stetler WR; Department of Neurological Surgery, University of Alabama at Birmingham, Birmingham, Alabama, USA.
  • Polifka AJ; Department of Neurological Surgery, University of Florida, Gainesville, Florida, USA.
  • Hoh BL; Department of Neurological Surgery, University of Florida, Gainesville, Florida, USA.
  • Arreola MM; Department of Radiology, University of Florida, Gainesville, Florida, USA.
  • Fox WC; Department of Neurological Surgery, University of Florida, Gainesville, Florida, USA.
World Neurosurg ; 110: e636-e641, 2018 Feb.
Article en En | MEDLINE | ID: mdl-29180081
ABSTRACT

BACKGROUND:

In some fluoroscopy machines, the dose-rate output of the fluoroscope is tied to a selectable patient size. Although patient size may play a significant role in visceral or cardiac procedures, head morphology is less variable, and high dose outputs may not be necessary even in very obese patients. We hypothesized that very small patient size setting can be used to reduce dose for cerebral angiography without compromising image quality.

METHODS:

Patients who underwent endovascular neurosurgical procedures during the 2015-2016 academic year were identified, and estimated procedural air kerma (AK) was tabulated retrospectively. Technologists were instructed to begin using the very small patient size setting for all procedures performed using our Philips Allura Xper FD20 biplane fluoroscopy system beginning in March 2016. No changes were made in a second procedure room using a Toshiba Infinix system. Student t tests and logistic regression models were used to compare radiation exposure before and after March 1, 2016, for both machines.

RESULTS:

For diagnostic cerebral angiograms performed on the Philips system (n = 302), AK was reduced by approximately 17% (1277 vs. 1061 mGy; P = 0.0006.) Changes in table height, total fluoroscopy time, patient weight, and body mass index did not contribute to this difference. No significant change was seen in total AK using the Toshiba system (n = 237). Blinded review by a neuroradiologist did not demonstrate any change in image quality.

CONCLUSIONS:

Using the very small patient size reduces fluoroscopy dose by 17% for cerebral angiography without impacting image quality.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Peso Corporal / Angiografía Cerebral / Fluoroscopía / Procedimientos Neuroquirúrgicos / Cirugía Asistida por Computador / Procedimientos Endovasculares Tipo de estudio: Observational_studies / Prognostic_studies Límite: Humans Idioma: En Revista: World Neurosurg Asunto de la revista: NEUROCIRURGIA Año: 2018 Tipo del documento: Article

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Peso Corporal / Angiografía Cerebral / Fluoroscopía / Procedimientos Neuroquirúrgicos / Cirugía Asistida por Computador / Procedimientos Endovasculares Tipo de estudio: Observational_studies / Prognostic_studies Límite: Humans Idioma: En Revista: World Neurosurg Asunto de la revista: NEUROCIRURGIA Año: 2018 Tipo del documento: Article