Your browser doesn't support javascript.
loading
An Embedded Device for Real-Time Noninvasive Intracranial Pressure Estimation.
Matthews, Jonathan M; Fanelli, Andrea; Heldt, Thomas.
Afiliación
  • Matthews JM; Department of Electrical Engineering and Computer Science, Massachusetts Institute of Technology, Cambridge, MA, USA.
  • Fanelli A; Institute for Medical Engineering and Science, Massachusetts Institute of Technology, Cambridge, MA, USA.
  • Heldt T; Institute for Medical Engineering and Science, Massachusetts Institute of Technology, Cambridge, MA, USA.
Acta Neurochir Suppl ; 126: 85-88, 2018.
Article en En | MEDLINE | ID: mdl-29492538
ABSTRACT

OBJECTIVE:

The monitoring of intracranial pressure (ICP) is indicated for diagnosing and guiding therapy in many neurological conditions. Current monitoring methods, however, are highly invasive, limiting their use to the most critically ill patients only. Our goal is to develop and test an embedded device that performs all necessary mathematical operations in real-time for noninvasive ICP (nICP) estimation based on a previously developed model-based approach that uses cerebral blood flow velocity (CBFV) and arterial blood pressure (ABP) waveforms. MATERIALS AND

METHODS:

The nICP estimation algorithm along with the required preprocessing steps were implemented on an NXP LPC4337 microcontroller unit (MCU). A prototype device using the MCU was also developed, complete with display, recording functionality, and peripheral interfaces for ABP and CBFV monitoring hardware.

RESULTS:

The device produces an estimate of mean ICP once per minute and performs the necessary computations in 410 ms, on average. Real-time nICP estimates differed from the original batch-mode MATLAB implementation of theestimation algorithm by 0.63 mmHg (root-mean-square error).

CONCLUSIONS:

We have demonstrated that real-time nICP estimation is possible on a microprocessor platform, which offers the advantages of low cost, small size, and product modularity over a general-purpose computer. These attributes take a step toward the goal of real-time nICP estimation at the patient's bedside in a variety of clinical settings.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Velocidad del Flujo Sanguíneo / Algoritmos / Presión Intracraneal / Circulación Cerebrovascular / Hipertensión Intracraneal / Equipos y Suministros / Presión Arterial Tipo de estudio: Prognostic_studies Límite: Humans Idioma: En Revista: Acta Neurochir Suppl Año: 2018 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Velocidad del Flujo Sanguíneo / Algoritmos / Presión Intracraneal / Circulación Cerebrovascular / Hipertensión Intracraneal / Equipos y Suministros / Presión Arterial Tipo de estudio: Prognostic_studies Límite: Humans Idioma: En Revista: Acta Neurochir Suppl Año: 2018 Tipo del documento: Article País de afiliación: Estados Unidos