Comparison of strain and acoustic radiation force impulse elastography of breast lesions by qualitative evaluation.
Zhao, Qing; Wang, Xiao-Lei; Sun, Jia-Wei; Jiang, Zhao-Peng; Tao, Lin; Zhou, Xian-Li.
Afiliación
  • Zhao Q; In-Patient Ultrasound Department, Second Affiliated Hospital of Harbin Medical University, Nan Gang, Harbin, Heilongjiang, China.
  • Wang XL; In-Patient Ultrasound Department, Second Affiliated Hospital of Harbin Medical University, Nan Gang, Harbin, Heilongjiang, China.
  • Sun JW; In-Patient Ultrasound Department, Second Affiliated Hospital of Harbin Medical University, Nan Gang, Harbin, Heilongjiang, China.
  • Jiang ZP; In-Patient Ultrasound Department, Second Affiliated Hospital of Harbin Medical University, Nan Gang, Harbin, Heilongjiang, China.
  • Tao L; In-Patient Ultrasound Department, Second Affiliated Hospital of Harbin Medical University, Nan Gang, Harbin, Heilongjiang, China.
  • Zhou XL; In-Patient Ultrasound Department, Second Affiliated Hospital of Harbin Medical University, Nan Gang, Harbin, Heilongjiang, China.
Clin Hemorheol Microcirc ; 70(1): 39-50, 2018.
Article en En | MEDLINE | ID: mdl-29660916
ABSTRACT

OBJECTIVES:

To compare the diagnostic performance of conventional strain elastography (CSE) and acoustic radiation force impulse (ARFI) induced SE for qualitative assessment of breast lesions and evaluate the additional value of the two techniques combined with Breast Imaging Reporting and Data System (BI-RADS) respectively for the differentiation of benign and malignant breast lesions.

METHODS:

In a cohort of 110 women, the conventional ultrasound (US) features and the elasticity scores of CSE and ARFI induced SE were recorded. The diagnostic performances of BI-RADS, elastography and BI-RADS plus elastography were evaluated, including the area under the receiver operating characteristic curve (AUROC), sensitivity, specificity and accuracy.

RESULTS:

Pathologically, there were forty-eight malignant and sixty-two benign breast lesions in the final analysis. The AUCs for CSE and ARFI induced SE are similar (CSE, 0.807; ARFI induced SE, 0.846; p > 0.05), however, the specificity of the latter method was significantly higher than that of CSE (83.9% vs. 58.1%, p = 0.004) in differentiating breast lesions. The accuracy and specificity of BI-RADS plus ARFI induced SE (84.5%, 80.6%, respectively) were significantly higher than BI-RADS alone (73.6%, 54.8%, respectively) and BI-RADS plus conventional SE (72.7%, 56.5%, respectively), respectively (p < 0.05) without loss of sensitivity.

CONCLUSIONS:

Our study showed that BI-RADS plus ARFI induced SE had a better diagnostic performance in the diagnosis of breast lesions in comparison with BI-RADS alone or BI-RADS plus CSE.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Neoplasias de la Mama / Ultrasonografía Mamaria / Ultrasonografía / Diagnóstico por Imagen de Elasticidad Tipo de estudio: Diagnostic_studies / Observational_studies / Qualitative_research Límite: Female / Humans / Middle aged Idioma: En Revista: Clin Hemorheol Microcirc Asunto de la revista: ANGIOLOGIA / HEMATOLOGIA Año: 2018 Tipo del documento: Article País de afiliación: China

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Neoplasias de la Mama / Ultrasonografía Mamaria / Ultrasonografía / Diagnóstico por Imagen de Elasticidad Tipo de estudio: Diagnostic_studies / Observational_studies / Qualitative_research Límite: Female / Humans / Middle aged Idioma: En Revista: Clin Hemorheol Microcirc Asunto de la revista: ANGIOLOGIA / HEMATOLOGIA Año: 2018 Tipo del documento: Article País de afiliación: China