Dual-energy CT in patients with abdominal malignant lymphoma: impact of noise-optimised virtual monoenergetic imaging on objective and subjective image quality.
Lenga, L; Czwikla, R; Wichmann, J L; Leithner, D; Albrecht, M H; D'Angelo, T; Arendt, C T; Booz, C; Hammerstingl, R; Vogl, T J; Martin, S S.
Afiliación
  • Lenga L; Department of Diagnostic and Interventional Radiology, University Hospital Frankfurt, Frankfurt, Germany.
  • Czwikla R; Department of Diagnostic and Interventional Radiology, University Hospital Frankfurt, Frankfurt, Germany.
  • Wichmann JL; Department of Diagnostic and Interventional Radiology, University Hospital Frankfurt, Frankfurt, Germany. Electronic address: docwichmann@gmail.com.
  • Leithner D; Department of Diagnostic and Interventional Radiology, University Hospital Frankfurt, Frankfurt, Germany.
  • Albrecht MH; Department of Diagnostic and Interventional Radiology, University Hospital Frankfurt, Frankfurt, Germany; Department of Radiology and Radiological Science, Medical University of South Carolina, Charleston, SC, USA.
  • D'Angelo T; Department of Diagnostic and Interventional Radiology, University Hospital Frankfurt, Frankfurt, Germany; Department of Biomedical Sciences and Morphological and Functional Imaging, University Hospital Messina, Messina, Italy.
  • Arendt CT; Department of Diagnostic and Interventional Radiology, University Hospital Frankfurt, Frankfurt, Germany.
  • Booz C; Department of Diagnostic and Interventional Radiology, University Hospital Frankfurt, Frankfurt, Germany.
  • Hammerstingl R; Department of Diagnostic and Interventional Radiology, University Hospital Frankfurt, Frankfurt, Germany.
  • Vogl TJ; Department of Diagnostic and Interventional Radiology, University Hospital Frankfurt, Frankfurt, Germany.
  • Martin SS; Department of Diagnostic and Interventional Radiology, University Hospital Frankfurt, Frankfurt, Germany.
Clin Radiol ; 73(9): 833.e19-833.e27, 2018 09.
Article en En | MEDLINE | ID: mdl-29884524
ABSTRACT

AIM:

To investigate the impact of noise-optimised virtual monoenergetic imaging (VMI+) reconstructions on quantitative and qualitative image parameters in patients with malignant lymphoma at dual-energy computed tomography (DECT) examinations of the abdomen. MATERIALS AND

METHODS:

Thirty-five consecutive patients (mean age, 53.8±18.6 years; range, 21-82 years) with histologically proven malignant lymphoma of the abdomen were included retrospectively. Images were post-processed with standard linear blending (M_0.6), traditional VMI, and VMI+ technique at energy levels ranging from 40 to 100 keV in 10 keV increments. Signal-to-noise (SNR) and contrast-to-noise ratios (CNR) were objectively measured in lymphoma lesions. Image quality, lesion delineation, and image noise were rated subjectively by three blinded observers using five-point Likert scales.

RESULTS:

Quantitative image quality parameters peaked at 40-keV VMI+ (SNR, 15.77±7.74; CNR, 18.27±8.04) with significant differences compared to standard linearly blended M_0.6 (SNR, 7.96±3.26; CNR, 13.55±3.47) and all traditional VMI series (p<0.001). Qualitative image quality assessment revealed significantly superior ratings for image quality at 60-keV VMI+ (median, 5) in comparison with all other image series (p<0.001). Assessment of lesion delineation showed the highest rating scores for 40-keV VMI+ series (median, 5), while lowest subjective image noise was found for 100-keV VMI+ reconstructions (median, 5).

CONCLUSION:

Low-keV VMI+ reconstructions led to improved image quality and lesion delineation of malignant lymphoma lesions compared to standard image reconstruction and traditional VMI at abdominal DECT examinations.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Interpretación de Imagen Radiográfica Asistida por Computador / Tomografía Computarizada por Rayos X / Imagen Radiográfica por Emisión de Doble Fotón / Abdomen / Linfoma Tipo de estudio: Observational_studies / Qualitative_research Límite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Clin Radiol Año: 2018 Tipo del documento: Article País de afiliación: Alemania

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Interpretación de Imagen Radiográfica Asistida por Computador / Tomografía Computarizada por Rayos X / Imagen Radiográfica por Emisión de Doble Fotón / Abdomen / Linfoma Tipo de estudio: Observational_studies / Qualitative_research Límite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Clin Radiol Año: 2018 Tipo del documento: Article País de afiliación: Alemania