Your browser doesn't support javascript.
loading
Clinical Validation of a Novel High-Sensitivity Cardiac Troponin I Assay for Early Diagnosis of Acute Myocardial Infarction.
Boeddinghaus, Jasper; Twerenbold, Raphael; Nestelberger, Thomas; Badertscher, Patrick; Wildi, Karin; Puelacher, Christian; du Fay de Lavallaz, Jeanne; Keser, Elif; Rubini Giménez, Maria; Wussler, Desiree; Kozhuharov, Nikola; Rentsch, Katharina; Miró, Òscar; Martin-Sanchez, F Javier; Morawiec, Beata; Stefanelli, Sabrina; Geigy, Nicolas; Keller, Dagmar I; Reichlin, Tobias; Mueller, Christian.
Afiliación
  • Boeddinghaus J; Cardiovascular Research Institute Basel (CRIB) and Department of Cardiology, University Hospital Basel, University of Basel, Basel, Switzerland.
  • Twerenbold R; Division of Internal Medicine, University Hospital Basel, University of Basel, Basel, Switzerland.
  • Nestelberger T; GREAT network, Rome, Italy.
  • Badertscher P; Cardiovascular Research Institute Basel (CRIB) and Department of Cardiology, University Hospital Basel, University of Basel, Basel, Switzerland.
  • Wildi K; GREAT network, Rome, Italy.
  • Puelacher C; Department of General and Interventional Cardiology, Hamburg University Heart Center, Hamburg, Germany.
  • du Fay de Lavallaz J; Cardiovascular Research Institute Basel (CRIB) and Department of Cardiology, University Hospital Basel, University of Basel, Basel, Switzerland.
  • Keser E; GREAT network, Rome, Italy.
  • Rubini Giménez M; Cardiovascular Research Institute Basel (CRIB) and Department of Cardiology, University Hospital Basel, University of Basel, Basel, Switzerland.
  • Wussler D; GREAT network, Rome, Italy.
  • Kozhuharov N; Cardiovascular Research Institute Basel (CRIB) and Department of Cardiology, University Hospital Basel, University of Basel, Basel, Switzerland.
  • Rentsch K; GREAT network, Rome, Italy.
  • Miró Ò; Cardiovascular Research Institute Basel (CRIB) and Department of Cardiology, University Hospital Basel, University of Basel, Basel, Switzerland.
  • Martin-Sanchez FJ; GREAT network, Rome, Italy.
  • Morawiec B; Cardiovascular Research Institute Basel (CRIB) and Department of Cardiology, University Hospital Basel, University of Basel, Basel, Switzerland.
  • Stefanelli S; Division of Internal Medicine, University Hospital Basel, University of Basel, Basel, Switzerland.
  • Geigy N; GREAT network, Rome, Italy.
  • Keller DI; Cardiovascular Research Institute Basel (CRIB) and Department of Cardiology, University Hospital Basel, University of Basel, Basel, Switzerland.
  • Reichlin T; GREAT network, Rome, Italy.
  • Mueller C; Cardiovascular Research Institute Basel (CRIB) and Department of Cardiology, University Hospital Basel, University of Basel, Basel, Switzerland.
Clin Chem ; 64(9): 1347-1360, 2018 09.
Article en En | MEDLINE | ID: mdl-29941469
ABSTRACT

BACKGROUND:

Clinical performance of the novel high-sensitivity cardiac troponin I (Siemens-hs-cTnI-Centaur) assay is unknown. We aimed to clinically validate the Siemens-hs-cTnI-Centaur assay and develop 0/1-h and 0/2-h algorithms.

METHODS:

We enrolled patients presenting to the emergency department with symptoms suggestive of acute myocardial infarction (AMI). Final diagnoses were centrally adjudicated by 2 independent cardiologists including all clinical information twice first, using serial hs-cTnT (Roche-Elecsys, primary analysis); second, using hs-cTnI (Abbott-Architect, secondary analysis) measurements in addition to the clinically applied (hs)-cTn. Siemens-hs-cTnI-Centaur was measured at presentation, 1 h, and 2 h. The primary objective was a direct comparison of diagnostic accuracy, quantified by the area under the ROC curve (AUC), of Siemens-hs-cTnI-Centaur vs the 2 established hs-cTn assays (Roche-hs-cTnT-Elecsys, Abbott-hs-cTnI-Architect). Secondary objectives included the development of Siemens-hs-cTnI-Centaur-specific 0/1-h and 0/2-h algorithms.

RESULTS:

AMI was the final diagnosis in 318 of 1755 (18%) patients (using Roche-hs-cTnT-Elecsys for adjudication). The AUC at presentation for Siemens-hs-cTnI-Centaur was 0.94 (95% CI, 0.92-0.96) and comparable with 0.95 (95% CI, 0.93-0.97) for Roche-hs-cTnT-Elecsys and 0.93 (95% CI, 0.90-0.96) for Abbott-hs-cTnI-Architect. Applying the derived Siemens-hs-cTnI-Centaur 0/1-h algorithm to the validation cohort, 46% of patients were ruled out (sensitivity, 99.1%; 95% CI, 95.3-100), and 18% of patients were ruled in (specificity, 94.1%; 95% CI, 91.8-95.9). The Siemens-hs-cTnI-Centaur 0/2-h algorithm ruled out 55% of patients (sensitivity, 100%; 95% CI, 94.1-100), and ruled in 18% of patients (specificity, 96.0%; 95% CI, 93.1-97.9). Findings were confirmed in the secondary analyses using serial measurements of Abbott-hs-cTnI-Architect for adjudication.

CONCLUSIONS:

Diagnostic accuracy and clinical utility of the novel Siemens-hs-cTnI-Centaur assay are high and comparable with the established hs-cTn assays. ClinicalTrials.gov Identifier NCT00470587.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Troponina I / Infarto del Miocardio Tipo de estudio: Diagnostic_studies / Prognostic_studies / Screening_studies Límite: Female / Humans / Male Idioma: En Revista: Clin Chem Asunto de la revista: QUIMICA CLINICA Año: 2018 Tipo del documento: Article País de afiliación: Suiza

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Troponina I / Infarto del Miocardio Tipo de estudio: Diagnostic_studies / Prognostic_studies / Screening_studies Límite: Female / Humans / Male Idioma: En Revista: Clin Chem Asunto de la revista: QUIMICA CLINICA Año: 2018 Tipo del documento: Article País de afiliación: Suiza