Interference of Thalassemias on Hemoglobin A1c Measurement by Ion-Exchange High-Performance Liquid Chromatography Method Tosoh HLC-723 G8.
Clin Lab ; 64(7): 1171-1175, 2018 Jul 01.
Article en En | MEDLINE | ID: mdl-30146829
ABSTRACT

BACKGROUND:

The presence of hemoglobinopathies could interfere with some assays for Hemoglobin A1c (HbA1c) measurement; therefore, the effect of thalassemia on ion-exchange high-performance liquid chromatography (IEHPLC) method Tosoh HLC-723 G8 (Tosoh G8) was evaluated.

METHODS:

A total of 43 normal controls and 101 thalassemia patients were quantified by Premier Hb9210 and Tosoh G8 (variant-mode) systems. At the same time, 7 normal controls and 8 thalassemia patients were confirmed by the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) reference method for verification.

RESULTS:

For normal controls, the HbA1c values of Tosoh G8 system (y) showed great correlation and agreement with those from Premier Hb9210 (x) (y = 0.9688x + 0.2151, r = 0.9951; mean difference 0.02 ± 0.30%), and no significant relative bias above 7% was observed; the HbA1c values obtained by Tosoh G8 were consistent with the IFCC targets (relative bias < ± 6%) in all of the samples. However, for thalassemia, the correlation between Tosoh G8 (y) and Premier Hb9210 (x) became relatively low (y = 0.8079x + 1.2897, r = 0.7780); the HbA1c values of 91.1% of the samples (92/101) obtained by Tosoh G8 were higher than those by Premier Hb9210 (mean difference 0.33 ± 0.48%) and a significant positive bias above 7% was noticed in 43.3% (45/101) thalassemia patients; when compared with the IFCC targets, the 87.5% (7/8) relative bias was > ± 6%.

CONCLUSIONS:

Thalassemia could directly affect the measurement of HbA1c using the IE-HPLC method Tosoh G8 and the clinical laboratorial staff should pay close attention.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Talasemia / Hemoglobina Glucada / Cromatografía Líquida de Alta Presión / Técnicas de Laboratorio Clínico / Servicios de Laboratorio Clínico Tipo de estudio: Diagnostic_studies Límite: Adult / Humans Idioma: En Revista: Clin Lab Asunto de la revista: TECNICAS E PROCEDIMENTOS DE LABORATORIO Año: 2018 Tipo del documento: Article

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Talasemia / Hemoglobina Glucada / Cromatografía Líquida de Alta Presión / Técnicas de Laboratorio Clínico / Servicios de Laboratorio Clínico Tipo de estudio: Diagnostic_studies Límite: Adult / Humans Idioma: En Revista: Clin Lab Asunto de la revista: TECNICAS E PROCEDIMENTOS DE LABORATORIO Año: 2018 Tipo del documento: Article