Usability Study of the User-Interface of Intensive Care Ventilators Based on User Test and Eye-Tracking Signals.
Jiang, Mingyin; Liu, Shenglin; Feng, Qingmin; Gao, Jiaqi; Zhang, Qiang.
Afiliación
  • Jiang M; Department of Medical Engineering, Union Hospital, Tongji Medical College, Huazhong University of Science and Technology, Wuhan, Hubei, China (mainland).
  • Liu S; Healthcare Ergonomics Lab, Union Hospital, Tongji Medical College, Huazhong University of Science and Technology, Wuhan, Hubei, China (mainland).
  • Feng Q; Department of Medical Engineering, Union Hospital, Tongji Medical College, Huazhong University of Science and Technology, Wuhan, Hubei, China (mainland).
  • Gao J; Healthcare Ergonomics Lab, Union Hospital, Tongji Medical College, Huazhong University of Science and Technology, Wuhan, Hubei, China (mainland).
  • Zhang Q; Department of Medical Engineering, Union Hospital, Tongji Medical College, Huazhong University of Science and Technology, Wuhan, Hubei, China (mainland).
Med Sci Monit ; 24: 6617-6629, 2018 Sep 20.
Article en En | MEDLINE | ID: mdl-30232319
ABSTRACT
BACKGROUND The aim of this study was to evaluate the ergonomics of the user-interface for 3 intensive care ventilators, and identify usability problems leading to user errors. MATERIAL AND METHODS Sixteen respiratory therapists were recruited to perform 6 specific user-interface operational tasks on ventilators. Data (task completion time, pupil diameter, average slope of pupil diameter change, and subjective evaluation) were collected through objective measurement, questionnaires, and an eye-tracking instrument. RESULTS For task completion time, there were significant differences among ventilators in recognition tasks of ventilator mode and settings (P<0.05), modification of ventilator modes and recognizing (P<0.05) and changing alarm settings (P<0.05). A mean of 15±2 task failures was observed for each ventilator. For the change of pupil diameter, a significant difference was observed between ventilators (except task 2, P<0.05). For average slope of pupil diameter change, a significant difference was also observed between ventilators (except task 2, P<0.05). The Servo I showed a better correlation between task completion time and pupil diameter change. The subjective evaluation results were clear Evital 4 received worst scores in terms of friendliness of user-interface, information display and safety (respectively, P<0.05). CONCLUSIONS The present study provided valuable evidence to indicate the ergonomic of ventilators now used in China. With the result of this study, we can infer that the Evital 4 were poorly ergonomic designed. Furthermore, the study also demonstrated that eye-tracking can be a promising tool to evaluate the ergonomics of the user-interface.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Respiración Artificial / Interfaz Usuario-Computador / Ergonomía Tipo de estudio: Prognostic_studies Límite: Humans País/Región como asunto: Asia Idioma: En Revista: Med Sci Monit Asunto de la revista: MEDICINA Año: 2018 Tipo del documento: Article

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Respiración Artificial / Interfaz Usuario-Computador / Ergonomía Tipo de estudio: Prognostic_studies Límite: Humans País/Región como asunto: Asia Idioma: En Revista: Med Sci Monit Asunto de la revista: MEDICINA Año: 2018 Tipo del documento: Article