Assessment of subjective refraction with a clinical adaptive optics visual simulator.
Hervella, Lucía; Villegas, Eloy A; Prieto, Pedro M; Artal, Pablo.
Afiliación
  • Hervella L; Laboratorio de Óptica, Universidad de Murcia, Murcia, Spain; Departamento de Física, Universidad de Murcia, and Voptica SL, Murcia, Spain.
  • Villegas EA; Laboratorio de Óptica, Universidad de Murcia, Murcia, Spain. Electronic address: villegas@um.es.
  • Prieto PM; Laboratorio de Óptica, Universidad de Murcia, Murcia, Spain.
  • Artal P; Laboratorio de Óptica, Universidad de Murcia, Murcia, Spain.
J Cataract Refract Surg ; 45(1): 87-93, 2019 01.
Article en En | MEDLINE | ID: mdl-30309774
ABSTRACT

PURPOSE:

To clinically validate an adaptive optics visual simulator (VAO) that measures subjective refraction and visual acuity.

SETTING:

Optics Laboratory, University of Murcia, Murcia, Spain.

DESIGN:

Prospective case series.

METHODS:

Using the adaptive optics visual simulator, 2 examiners measured the subjective refraction and visual acuity in healthy eyes of volunteers; 1 examiner also used a trial frame as a gold standard. The interexaminer reproducibility and agreement with the gold standard were estimated using the following statistical parameters limits of agreement from Bland-Altman analysis, significance between differences (P value), and intraclass correlation coefficient (ICC).

RESULTS:

Seventy-six eyes of 38 volunteers were measured. Interexaminer reproducibility for subjective refraction was excellent (ICC ≥0.96; P > .05), with low 95% confidence interval (CI) values for the power vectors M (spherical equivalent of the given refractive error), J0 (Jackson cross-cylinder, axes at 180 degrees and 90 degrees), and J45 (Jackson cross-cylinder, axes at 45 degrees and 135 degrees) (±0.51 diopter [D], ±0.14 D, and ±0.14 D, respectively). No significant differences in subjective refraction and visual acuity were found between the visual simulator and gold standard (P > .05), with 95% CIs for M, J0, and J45 (subjective refraction) of ±0.67 D, ±0.14 D, and ±0.16 D, respectively, and a ±0.10 logarithm of the minimum angle of resolution (visual acuity).

CONCLUSION:

Subjective refraction results using the adaptive optics visual simulator agreed with those of the gold standard and can be used as the baseline for visual simulation of any optical corneal profile or intraocular lens design for refractive surgery patients.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Refracción Ocular / Agudeza Visual / Córnea / Aberrometría Tipo de estudio: Observational_studies Límite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: J Cataract Refract Surg Asunto de la revista: OFTALMOLOGIA Año: 2019 Tipo del documento: Article País de afiliación: España

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Refracción Ocular / Agudeza Visual / Córnea / Aberrometría Tipo de estudio: Observational_studies Límite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: J Cataract Refract Surg Asunto de la revista: OFTALMOLOGIA Año: 2019 Tipo del documento: Article País de afiliación: España