Pre-procedural fit-testing of TAVR valves using parametric modeling and 3D printing.
Hosny, Ahmed; Dilley, Joshua D; Kelil, Tatiana; Mathur, Moses; Dean, Mason N; Weaver, James C; Ripley, Beth.
Afiliación
  • Hosny A; Dana-Farber Cancer Institute, Harvard Medical School, USA. Electronic address: ahmed_hosny@dfci.harvard.edu.
  • Dilley JD; Massachusetts General Hospital, Department of Anesthesia, Critical Care & Pain Medicine, Harvard Medical School, USA. Electronic address: joshuaddilley@gmail.com.
  • Kelil T; Department of Radiology, Brigham and Women's Hospital, Harvard Medical School, USA. Electronic address: tatiana.kelil3d@gmail.com.
  • Mathur M; Department of Medicine, Division of Cardiology, University of Washington Medical Center, USA. Electronic address: moses.mathur@gmail.com.
  • Dean MN; Department of Biomaterials, Max Planck Institute of Colloids and Interfaces, Germany. Electronic address: mason.dean@mpikg.mpg.de.
  • Weaver JC; Wyss Institute for Biologically Inspired Engineering, Harvard University, USA. Electronic address: james.weaver@wyss.harvard.edu.
  • Ripley B; Department of Radiology, University of Washington, USA; VA Puget Sound Health Care System, USA. Electronic address: beth.ripley2@va.gov.
J Cardiovasc Comput Tomogr ; 13(1): 21-30, 2019.
Article en En | MEDLINE | ID: mdl-30322772
ABSTRACT

BACKGROUND:

Successful transcatheter aortic valve replacement (TAVR) requires an understanding of how a prosthetic valve will interact with a patient's anatomy in advance of surgical deployment. To improve this understanding, we developed a benchtop workflow that allows for testing of physical interactions between prosthetic valves and patient-specific aortic root anatomy, including calcified leaflets, prior to actual prosthetic valve placement.

METHODS:

This was a retrospective study of 30 patients who underwent TAVR at a single high volume center. By design, the dataset contained 15 patients with a successful annular seal (defined by an absence of paravalvular leaks) and 15 patients with a sub-optimal seal (presence of paravalvular leaks) on post-procedure transthoracic echocardiogram (TTE). Patients received either a balloon-expandable (Edwards Sapien or Sapien XT, n = 15), or a self-expanding (Medtronic CoreValve or Core Evolut, n = 14, St. Jude Portico, n = 1) valve. Pre-procedural computed tomography (CT) angiograms, parametric geometry modeling, and multi-material 3D printing were utilized to create flexible aortic root physical models, including displaceable calcified valve leaflets. A 3D printed adjustable sizing device was then positioned in the aortic root models and sequentially opened to larger valve sizes, progressively flattening the calcified leaflets against the aortic wall. Optimal valve size and fit were determined by visual inspection and quantitative pressure mapping of interactions between the sizer and models.

RESULTS:

Benchtop-predicted "best fit" valve size showed a statistically significant correlation with gold standard CT measurements of the average annulus diameter (n = 30, p < 0.0001 Wilcoxon matched-pairs signed rank test). Adequateness of seal (presence or absence of paravalvular leak) was correctly predicted in 11/15 (73.3%) patients who received a balloon-expandable valve, and in 9/15 (60%) patients who received a self-expanding valve. Pressure testing provided a physical map of areas with an inadequate seal; these corresponded to areas of paravalvular leak documented by post-procedural transthoracic echocardiography.

CONCLUSION:

We present and demonstrate the potential of a workflow for determining optimal prosthetic valve size that accounts for aortic annular dimensions as well as the active displacement of calcified valve leaflets during prosthetic valve deployment. The workflow's open source framework offers a platform for providing predictive insights into the design and testing of future prosthetic valves.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Válvula Aórtica / Estenosis de la Válvula Aórtica / Diseño de Prótesis / Calcinosis / Prótesis Valvulares Cardíacas / Reemplazo de la Válvula Aórtica Transcatéter / Modelación Específica para el Paciente / Impresión Tridimensional Tipo de estudio: Diagnostic_studies / Etiology_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Límite: Aged / Aged80 / Female / Humans / Male Idioma: En Revista: J Cardiovasc Comput Tomogr Asunto de la revista: ANGIOLOGIA / CARDIOLOGIA / RADIOLOGIA Año: 2019 Tipo del documento: Article

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Válvula Aórtica / Estenosis de la Válvula Aórtica / Diseño de Prótesis / Calcinosis / Prótesis Valvulares Cardíacas / Reemplazo de la Válvula Aórtica Transcatéter / Modelación Específica para el Paciente / Impresión Tridimensional Tipo de estudio: Diagnostic_studies / Etiology_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Límite: Aged / Aged80 / Female / Humans / Male Idioma: En Revista: J Cardiovasc Comput Tomogr Asunto de la revista: ANGIOLOGIA / CARDIOLOGIA / RADIOLOGIA Año: 2019 Tipo del documento: Article