The relationship between seizure in electroconvulsive therapy and pupillary response using an automated pupilometer.
Shirozu, Kazuhiro; Murayama, Keitaro; Karashima, Yuji; Setoguchi, Hidekazu; Miura, Tomofumi; Hoka, Sumio.
Afiliación
  • Shirozu K; Department of Anesthesiology and Critical Care Medicine, Kyushu University Hospital, Kyushu University, 3-1-1 Maidashi, Higashi-ku, Fukuoka, 812-8582, Japan. shiron@kuaccm.med.kyushu-u.ac.jp.
  • Murayama K; Department of Neuropsychiatry, Kyushu University Hospital, Fukuoka, Japan.
  • Karashima Y; Department of Anesthesiology and Critical Care Medicine, Kyushu University Graduate School of Medicine, Fukuoka, Japan.
  • Setoguchi H; Operating Rooms, Kyushu University Hospital, Fukuoka, Japan.
  • Miura T; Department of Neuro-Psychiatry, Kyushu University Graduate School of Medicine, Fukuoka, Japan.
  • Hoka S; Department of Anesthesiology and Critical Care Medicine, Kyushu University Graduate School of Medicine, Fukuoka, Japan.
J Anesth ; 32(6): 866-871, 2018 12.
Article en En | MEDLINE | ID: mdl-30327871
ABSTRACT

OBJECTIVES:

Seizure duration and morphology, postictal suppression, and sympathetic nervous system activation are all recommended as assessments of adequate seizure in electroconvulsive therapy (ECT). However, blood pressure and heart rate are not typically assessed as part of sympathetic nervous system activation because of the administration of anesthetic or cardiovascular agents during ECT. Although the pupils are known to reflect to the activity of autonomic nervous system and the degree of brain damage, previous studies have not examined the relationship between seizure of electroconvulsive therapy and pupillary response.

METHODS:

We conducted 98 sessions of ECT with 13 patients, divided into two groups according to seizure quality (1) adequate or (2) inadequate. Pupillary light reflex [% constriction = (maximum resting pupil size {MAX} - minimum pupil size after light stimulation)/MAX × 100] was measured using a portable infrared quantitative pupilometer before anesthesia induction and immediately after electrical stimulation.

RESULTS:

The number regarded as adequate was 67 times and as inadequate was 31 times. Maximum pupil size at the control and immediately after electrical stimulation was similar between the adequate and inadequate groups. Pupillary light reflex was similar at the control between both groups, but significantly smaller immediately after stimulation in the adequate group (2.5 ± 3.6%) compared with the inadequate group (10.6 ± 11.5%). Receiver operating characteristic curve analysis revealed that pupillary light reflex (> 5.5%) predicted adequate seizure.

CONCLUSIONS:

The current findings suggest that pupillary constriction immediately after ECT could provide a helpful method for assessing the efficacy of ECT.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Reflejo Pupilar / Convulsiones / Terapia Electroconvulsiva Tipo de estudio: Prognostic_studies Límite: Adult / Aged / Aged80 / Humans / Middle aged Idioma: En Revista: J Anesth Asunto de la revista: ANESTESIOLOGIA Año: 2018 Tipo del documento: Article País de afiliación: Japón

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Reflejo Pupilar / Convulsiones / Terapia Electroconvulsiva Tipo de estudio: Prognostic_studies Límite: Adult / Aged / Aged80 / Humans / Middle aged Idioma: En Revista: J Anesth Asunto de la revista: ANESTESIOLOGIA Año: 2018 Tipo del documento: Article País de afiliación: Japón