TYPE TESTING OF LiF:Mg,Cu,P (TLD-100H) WHOLE-BODY DOSEMETERS FOR THE ASSESSMENT OF Hp(10) AND Hp(0.07).
Pereira, J S; Pereira, M F; Rangel, S; Caldeira, M; Carvalhal, G; Santos, L M; Cardoso, J V; Alves, J G.
Afiliación
  • Pereira JS; Universidade de Lisboa (UL), Instituto Superior Técnico (IST), Laboratório de Proteção e Segurança Radiológica (LPSR), Estrada Nacional 10 (ao km 139,7), Bobadela LRS, Portugal.
  • Pereira MF; UL-IST, Centro de Ciências e Tecnologias Nucleares (C2TN), Estrada Nacional 10 (ao km 139,7), Bobadela LRS, Portugal.
  • Rangel S; Universidade de Lisboa (UL), Instituto Superior Técnico (IST), Laboratório de Proteção e Segurança Radiológica (LPSR), Estrada Nacional 10 (ao km 139,7), Bobadela LRS, Portugal.
  • Caldeira M; UL-IST, Centro de Ciências e Tecnologias Nucleares (C2TN), Estrada Nacional 10 (ao km 139,7), Bobadela LRS, Portugal.
  • Carvalhal G; Universidade de Lisboa (UL), Instituto Superior Técnico (IST), Laboratório de Proteção e Segurança Radiológica (LPSR), Estrada Nacional 10 (ao km 139,7), Bobadela LRS, Portugal.
  • Santos LM; Universidade de Lisboa (UL), Instituto Superior Técnico (IST), Laboratório de Proteção e Segurança Radiológica (LPSR), Estrada Nacional 10 (ao km 139,7), Bobadela LRS, Portugal.
  • Cardoso JV; UL-IST, Centro de Ciências e Tecnologias Nucleares (C2TN), Estrada Nacional 10 (ao km 139,7), Bobadela LRS, Portugal.
  • Alves JG; Universidade de Lisboa (UL), Instituto Superior Técnico (IST), Laboratório de Proteção e Segurança Radiológica (LPSR), Estrada Nacional 10 (ao km 139,7), Bobadela LRS, Portugal.
Radiat Prot Dosimetry ; 184(2): 216-223, 2019 Aug 01.
Article en En | MEDLINE | ID: mdl-30496554
ABSTRACT
In this work, the initial results of the type testing of the LiFMg,Cu,P (TLD-100H) whole-body personal dosemeters are presented. An assessment of reproducibility, linearity of the response, the residual signal as a function of the dose, energy and angular dependence of the response was performed. In general, the dosemeters show good reproducibility for different dose values and a linear behaviour for a range between 0.1 and 300 mSv. The detection limits obtained are lower than 50 µSv. The system presents a good energy and angular response for different radiation qualities.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Fósforo / Dosimetría Termoluminiscente / Monitoreo de Radiación / Recuento Corporal Total / Compuestos de Litio / Cobre / Fluoruros / Magnesio Límite: Humans Idioma: En Revista: Radiat Prot Dosimetry Año: 2019 Tipo del documento: Article País de afiliación: Portugal

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Fósforo / Dosimetría Termoluminiscente / Monitoreo de Radiación / Recuento Corporal Total / Compuestos de Litio / Cobre / Fluoruros / Magnesio Límite: Humans Idioma: En Revista: Radiat Prot Dosimetry Año: 2019 Tipo del documento: Article País de afiliación: Portugal