Clinical evaluation of the repeatability of ocular aberrometry obtained with a new pyramid wavefront sensor.
Plaza-Puche, Ana B; Salerno, Liberdade C; Versaci, Francesco; Romero, Daniel; Alio, Jorge L.
Afiliación
  • Plaza-Puche AB; Vissum, Alicante, Spain.
  • Salerno LC; Vissum, Alicante, Spain.
  • Versaci F; R&D Department, Costruzione Strumenti Oftalmici (CSO), Florence, Italy.
  • Romero D; Division of Ophthalmology, Universidad Miguel Hernández, Alicante, Spain.
  • Alio JL; Vissum, Alicante, Spain.
Eur J Ophthalmol ; 29(6): 585-592, 2019 Nov.
Article en En | MEDLINE | ID: mdl-30516061
ABSTRACT

PURPOSE:

To evaluate the intrasubject repeatability of the ocular aberrometry obtained with a new ocular pyramidal aberrometer technology in a sample of normal eyes.

METHODS:

A total of 53 healthy eyes of 53 subjects with ages ranging from 18 to 45 years were included in this study. In all cases, three consecutive acquisitions were obtained. Intrasubject repeatability of the measurements with a pyramidal aberrometer was calculated. Intrasubject repeatability for 4.0- and 6.0-mm pupils was evaluated within the subject standard deviation (Sw) and intraclass correlation coefficient.

RESULTS:

Low values of the Sw and intraclass correlation coefficient outcomes close to 1 were observed for the sphere and cylinder at 3.0-mm pupil size. Most low Sw and intraclass correlation coefficient values close to 1 were observed for total, low-order aberrations and higher-order aberrations root mean square and for each Zernike coefficient analysis (intraclass correlation coefficient ⩾0.798) at 4.0-mm pupil size, with more limited outcomes for the aberrometric coefficient of Z(4, 4) with an intraclass correlation coefficient of 0.683. For a 6.0 mm pupil diameter, low Sw and intraclass correlation coefficient values close to 1 were observed for all aberrometric parameters or Zernike coefficients analyzed (intraclass correlation coefficient ⩾0.850).

CONCLUSION:

The new pyramidal aberrometer Osiris provides repeatable and consistent measurements of ocular aberrometry measurements in normal eyes.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Aberración de Frente de Onda Corneal / Aberrometría Tipo de estudio: Evaluation_studies / Observational_studies Límite: Adolescent / Adult / Female / Humans / Male / Middle aged Idioma: En Revista: Eur J Ophthalmol Asunto de la revista: OFTALMOLOGIA Año: 2019 Tipo del documento: Article País de afiliación: España

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Aberración de Frente de Onda Corneal / Aberrometría Tipo de estudio: Evaluation_studies / Observational_studies Límite: Adolescent / Adult / Female / Humans / Male / Middle aged Idioma: En Revista: Eur J Ophthalmol Asunto de la revista: OFTALMOLOGIA Año: 2019 Tipo del documento: Article País de afiliación: España