Technical Note: Relative proton stopping power estimation from virtual monoenergetic images reconstructed from dual-layer computed tomography.
Landry, Guillaume; Dörringer, Fabian; Si-Mohamed, Salim; Douek, Philippe; Abascal, Juan F P J; Peyrin, Françoise; Almeida, Isabel P; Verhaegen, Frank; Rinaldi, Ilaria; Parodi, Katia; Rit, Simon.
Afiliación
  • Landry G; Department of Medical Physics, Faculty of Physics, Ludwig-Maximilians-Universität München, Munich, Germany.
  • Dörringer F; Department of Medical Physics, Faculty of Physics, Ludwig-Maximilians-Universität München, Munich, Germany.
  • Si-Mohamed S; Radiology Department, Centre Hospitalier Universitaire, Lyon, France.
  • Douek P; Univ Lyon, INSA-Lyon, Université Claude Bernard Lyon 1, UJM-Saint-Étienne, CNRS, Inserm, CREATIS, UMR 5220, U1206, F-69373, Lyon, France.
  • Abascal JFPJ; Radiology Department, Centre Hospitalier Universitaire, Lyon, France.
  • Peyrin F; Univ Lyon, INSA-Lyon, Université Claude Bernard Lyon 1, UJM-Saint-Étienne, CNRS, Inserm, CREATIS, UMR 5220, U1206, F-69373, Lyon, France.
  • Almeida IP; Univ Lyon, INSA-Lyon, Université Claude Bernard Lyon 1, UJM-Saint-Étienne, CNRS, Inserm, CREATIS, UMR 5220, U1206, F-69373, Lyon, France.
  • Verhaegen F; Univ Lyon, INSA-Lyon, Université Claude Bernard Lyon 1, UJM-Saint-Étienne, CNRS, Inserm, CREATIS, UMR 5220, U1206, F-69373, Lyon, France.
  • Rinaldi I; ESRF, The European Synchrotron, Grenoble, France.
  • Parodi K; Department of Radiation Oncology (MAASTRO), GROW School for Oncology and Developmental Biology, Maastricht University Medical Centre, Maastricht, The Netherlands.
  • Rit S; Department of Radiation Oncology (MAASTRO), GROW School for Oncology and Developmental Biology, Maastricht University Medical Centre, Maastricht, The Netherlands.
Med Phys ; 46(4): 1821-1828, 2019 Apr.
Article en En | MEDLINE | ID: mdl-30695108
ABSTRACT

PURPOSE:

The objective of this technical note was to investigate the accuracy of proton stopping power relative to water (RSP) estimation using a novel dual-layer, dual-energy computed tomography (DL-DECT) scanner for potential use in proton therapy planning. DL-DECT allows dual-energy reconstruction from scans acquired at a single x-ray tube voltage V by using two-layered detectors.

METHODS:

Sets of calibration and evaluation inserts were scanned at a DL-DECT scanner in a custom phantom with variable diameter D (0 to 150 mm) at V of 120 and 140 kV. Inserts were additionally scanned at a synchrotron computed tomography facility to obtain comparative linear attenuation coefficients for energies from 50 to 100 keV, and reference RSP was obtained using a carbon ion beam and variable water column. DL-DECT monoenergetic (mono-E) reconstructions were employed to obtain RSP by adapting the Yang-Saito-Landry (YSL) method. The method was compared to reference RSP via the root mean square error (RMSE) over insert mean values obtained from volumetric regions of interest. The accuracy of intermediate quantities such as the relative electron density (RED), effective atomic number (EAN), and the mono-E was additionally evaluated.

RESULTS:

The lung inserts showed higher errors for all quantities and we report RMSE excluding them. RMSE for µ from DL-DECT mono-E was below 1.9%. For the evaluation inserts at D = 150 mm and V = 140 kV, RED RMSE was 1.0%, while for EAN it was 2.9%. RSP RMSE was below 0.8% for all D and V, which did not strongly affect the results.

CONCLUSIONS:

In this investigation of RSP accuracy from DL-DECT, we have shown that RMSE below 1% can be achieved. It was possible to adapt the YSL method for DL-DECT and intermediate quantities RED and EAN had comparable accuracy to previous publications.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Procesamiento de Imagen Asistido por Computador / Planificación de la Radioterapia Asistida por Computador / Tomografía Computarizada por Rayos X / Fantasmas de Imagen / Terapia de Protones / Pulmón / Neoplasias Tipo de estudio: Etiology_studies Límite: Humans Idioma: En Revista: Med Phys Año: 2019 Tipo del documento: Article País de afiliación: Alemania

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Procesamiento de Imagen Asistido por Computador / Planificación de la Radioterapia Asistida por Computador / Tomografía Computarizada por Rayos X / Fantasmas de Imagen / Terapia de Protones / Pulmón / Neoplasias Tipo de estudio: Etiology_studies Límite: Humans Idioma: En Revista: Med Phys Año: 2019 Tipo del documento: Article País de afiliación: Alemania