Impact of Novel Incorporation of CT-based Segment Mapping into a Conjugated Gradient Algorithm on Bone SPECT Imaging: Fundamental Characteristics of a Context-specific Reconstruction Method.
Okuda, Kyohei; Fujii, Susumu; Sakimoto, Shota.
Afiliación
  • Okuda K; Department of Clinical Radiology, Tottori University Hospital, Tottori, Japan.
  • Fujii S; Department of Clinical Radiology, Tottori University Hospital, Tottori, Japan.
  • Sakimoto S; Department of Clinical Radiology, Tottori University Hospital, Tottori, Japan.
Asia Ocean J Nucl Med Biol ; 7(1): 49-57, 2019.
Article en En | MEDLINE | ID: mdl-30705911
ABSTRACT

OBJECTIVES:

The latest single-photon emission computed tomography (SPECT)/computed tomography (CT) reconstruction system, referred to as xSPECT Bone™, is a context-specific reconstruction system utilizing tissue segmentation information from CT data, which is called a zone map. The aim of this study was to evaluate the effects of zone-map enhancement incorporated into the ordered-subset conjugated gradient minimization (OSCGM) reconstruction method on SPECT images.

METHODS:

Image quality with zone-map enhanced OSCGM (OSCGMz) and non-enhanced OSCGM methods was compared using various reconstruction parameters. The compartment phantom had 3 radioactive sections with CT values of about 1000, 250, and 0 HU. SPECT data were acquired using a low-energy high-resolution (LEHR) collimator, with a 256×256 matrix and 2.4-mm pixel size. The performances of the 2 reconstruction methods (OSCGM vs. OSCGMz) were evaluated on the basis of %error, coefficient of variation (%CV), and normalized mean squared error (NMSE), and adequate iterative update numbers were determined. The relative CV representing the ratio of smoothed images to non-smoothed images was calculated to evaluate the effects of the Gaussian filter on each section set with different CT values.

RESULTS:

On comparing the OSCGM and OSCGMz methods, it was found that the %error of the OSCGMz method tended to show convergence with fewer updates, especially in the high CT value section mimicking bone absorption. In the water section, the %CV of the OSCGMz method was lower than that of the OSCGM method. The NMSE minimum values for the OSCGM and OSCGMz methods were obtained at 30 and 20 updates, respectively. The relative CV for the OSCGMz method in the water section decreased remarkably according to the size of the full width at half maximum (FWHM) of the Gaussian filter.

CONCLUSION:

The zone-map enhancement contributed to SPECT reconstruction for the reproduction of radioactive concentrations in bone tissues, using a low number of OSCGM updates. Our findings indicated that the incorporation of zone maps into SPECT reconstruction might improve image quality.