No need for frame-wise attenuation correction in dynamic Rubidium-82 PET for myocardial blood flow quantification.
van Dijk, J D; Jager, P L; Ottervanger, J P; Slump, C H; van Dalen, J A.
Afiliación
  • van Dijk JD; Department of Nuclear Medicine, Isala Hospital, PO Box 10400, 8000 GK, Zwolle, The Netherlands. jorisvdijk@gmail.com.
  • Jager PL; Department of Nuclear Medicine, Isala Hospital, PO Box 10400, 8000 GK, Zwolle, The Netherlands.
  • Ottervanger JP; Department of Cardiology, Isala Hospital, PO Box 10400, 8000 GK, Zwolle, The Netherlands.
  • Slump CH; MIRA: Institute for Biomedical Technology and Technical Medicine, University of Twente, Enschede, The Netherlands.
  • van Dalen JA; Department of Medical Physics, Isala Hospital, PO Box 10400, 8000 GK, Zwolle, The Netherlands.
J Nucl Cardiol ; 26(3): 738-745, 2019 06.
Article en En | MEDLINE | ID: mdl-30790203
ABSTRACT

BACKGROUND:

Regadenoson-induced stress causes a repositioning of the heart, myocardial creep, in half of the patients undergoing Rubidium-82 (Rb-82) positron emission tomography (PET). As a result, misalignment of dynamic PET and computer tomography (CT) may occur, possibly affecting CT-based attenuation correction (AC) and thereby PET-based myocardial blood flow (MBF) quantification. Our aim was to determine the need for frame-wise PET-CT AC to obtain reliable MBF measurements.

METHODS:

31 Out of 64 consecutive patients had myocardial creep during regadenoson-induced stress Rb-82 PET-CT and were included. Prior to PET image reconstruction, we applied two AC methods; single PET-CT alignment and frame-wise alignment in which PET time-frames with myocardial creep were individually co-registered with CT. The PET-CT misalignment was then quantified and MBFs for the three vascular territories and whole myocardium were calculated and compared between both methods.

RESULTS:

The magnitude of misalignment due to myocardial creep was 13.8 ± 4.5 mm in caudal-cranial direction, 1.8 ± 2.1 mm in medial-lateral and 2.5 ± 1.8 mm in anterior-posterior direction. Frame-wise PET-CT registration did not result in different MBF measurements (P ≥ .07) and the magnitude of misalignment and MBF differences did not correlate (P ≥ .58).

CONCLUSION:

There is no need for frame-wise AC in dynamic Rb-82 PET for MBF quantification. Single alignment seems sufficient in patients with myocardial creep.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Enfermedad de la Arteria Coronaria / Radioisótopos de Rubidio / Imagen de Perfusión Miocárdica / Tomografía Computarizada por Tomografía de Emisión de Positrones / Movimientos de los Órganos Tipo de estudio: Observational_studies Límite: Aged / Female / Humans / Male / Middle aged Idioma: En Revista: J Nucl Cardiol Asunto de la revista: CARDIOLOGIA Año: 2019 Tipo del documento: Article País de afiliación: Países Bajos Pais de publicación: EEUU / ESTADOS UNIDOS / ESTADOS UNIDOS DA AMERICA / EUA / UNITED STATES / UNITED STATES OF AMERICA / US / USA

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Enfermedad de la Arteria Coronaria / Radioisótopos de Rubidio / Imagen de Perfusión Miocárdica / Tomografía Computarizada por Tomografía de Emisión de Positrones / Movimientos de los Órganos Tipo de estudio: Observational_studies Límite: Aged / Female / Humans / Male / Middle aged Idioma: En Revista: J Nucl Cardiol Asunto de la revista: CARDIOLOGIA Año: 2019 Tipo del documento: Article País de afiliación: Países Bajos Pais de publicación: EEUU / ESTADOS UNIDOS / ESTADOS UNIDOS DA AMERICA / EUA / UNITED STATES / UNITED STATES OF AMERICA / US / USA