Comparison of OA-2000 and IOL Master 500 using in cataract patients with high myopia.
Du, Ya-Li; Wang, Geng; Huang, Hui-Chun; Lin, Li-Yu; Jin, Chuang; Liu, Li-Fang; Liu, Xiu-Ru; Zhang, Ming-Zhi.
Afiliación
  • Du YL; The Joint Shantou International Eye Center of Shantou University and the Chinese University of Hong Kong, Shantou 515041, Guangdong Province, China.
  • Wang G; The Joint Shantou International Eye Center of Shantou University and the Chinese University of Hong Kong, Shantou 515041, Guangdong Province, China.
  • Huang HC; The Joint Shantou International Eye Center of Shantou University and the Chinese University of Hong Kong, Shantou 515041, Guangdong Province, China.
  • Lin LY; The Joint Shantou International Eye Center of Shantou University and the Chinese University of Hong Kong, Shantou 515041, Guangdong Province, China.
  • Jin C; The Joint Shantou International Eye Center of Shantou University and the Chinese University of Hong Kong, Shantou 515041, Guangdong Province, China.
  • Liu LF; The Joint Shantou International Eye Center of Shantou University and the Chinese University of Hong Kong, Shantou 515041, Guangdong Province, China.
  • Liu XR; The Joint Shantou International Eye Center of Shantou University and the Chinese University of Hong Kong, Shantou 515041, Guangdong Province, China.
  • Zhang MZ; The Joint Shantou International Eye Center of Shantou University and the Chinese University of Hong Kong, Shantou 515041, Guangdong Province, China.
Int J Ophthalmol ; 12(5): 844-847, 2019.
Article en En | MEDLINE | ID: mdl-31131247
ABSTRACT
This study was designed to compare optical biometry measurements and predicted refraction in cataract patients with high myopia [axial length (AL) ≥26 mm] using OA-2000 and IOL Master 500. Ocular biometry measurements were performed using both biometers before surgery. Uneventful cataract surgery was performed in all patients. Postoperative manifest refraction was obtained 3wk after surgery or later. A total of 67 eyes were examined. The AL, keratometry (K), and anterior chamber depth (ACD) of the two biometers showed excellent agreement. Predicted errors were similar and a strong positive correlation was observed (r=0.909). Out of 21 eyes (31.34%) with unsuccessful AL readings using the IOL Master 500, 20 eyes of them could be measured using OA-2000. Therefore, the biometric parameters measured by the two biometers showed good agreement. However, OA-2000 had a lower AL measurement failure rate.