A mechanistic investigation of the EDWARDS INTUITY Elite valve's hemodynamic performance.
Sadri, Vahid; Bloodworth, Charles H; Madukauwa-David, Immanuel David; Midha, Prem A; Raghav, Vrishank; Yoganathan, Ajit P.
Afiliación
  • Sadri V; Wallace H. Coulter Department of Biomedical Engineering, Georgia Institute of Technology and Emory University, Atlanta, GA, USA.
  • Bloodworth CH; Wallace H. Coulter Department of Biomedical Engineering, Georgia Institute of Technology and Emory University, Atlanta, GA, USA.
  • Madukauwa-David ID; George W. Woodruff School of Mechanical Engineering, Georgia Institute of Technology, Atlanta, GA, USA.
  • Midha PA; George W. Woodruff School of Mechanical Engineering, Georgia Institute of Technology, Atlanta, GA, USA.
  • Raghav V; Department of Aerospace Engineering, Auburn University, Auburn, AL, USA.
  • Yoganathan AP; Wallace H. Coulter Department of Biomedical Engineering, Georgia Institute of Technology and Emory University, Atlanta, GA, USA. ajit.yoganathan@bme.gatech.edu.
Gen Thorac Cardiovasc Surg ; 68(1): 9-17, 2020 Jan.
Article en En | MEDLINE | ID: mdl-31250203
ABSTRACT

OBJECTIVE:

Rapid deployment surgical aortic valve replacement has emerged as an alternative to the contemporary sutured valve technique. A difference in transvalvular pressure has been observed clinically between RD-SAVR and contemporary SAVR. A mechanistic inquiry into the impact of the rapid deployment valve inflow frame design on the left ventricular outflow tract and valve hemodynamics is needed.

METHODS:

A 23 mm EDWARDS INTUITY Elite rapid deployment valve and a control contemporary, sutured valve, a 23 mm Magna Ease valve, were implanted in an explanted human heart by an experienced cardiac surgeon. Per convention, the rapid deployment valve was implanted with three non-pledgeted, simple guiding sutures, while fifteen pledgeted, mattress sutures were used to implant the contemporary surgical valve. In vitro flow models were created from micro-computed tomography scans of the implanted valves and surrounding cardiac anatomy. Particle image velocimetry and hydrodynamic characterization experiments were conducted in the vicinity of the valves in a validated pulsatile flow loop system.

RESULTS:

The rapid deployment and control valves were found to have mean transvalvular pressure gradients of 7.92 ± 0.37 and 10.13 ± 0.48 mmHg, respectively. The inflow frame of the rapid deployment valve formed a larger, more circular, left ventricular outflow tract compared to the control valve. Furthermore, it was found that the presence of the control valve's sub-annular pledgets compromised its velocity distribution and consequently its pressure gradient.

CONCLUSIONS:

The rapid deployment valve's intra-annular inflow frame provides for a larger, left ventricular outflow tract, thus reducing the transvalvular pressure gradient and improving overall hemodynamic performance.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Prótesis Valvulares Cardíacas / Hemodinámica Límite: Humans Idioma: En Revista: Gen Thorac Cardiovasc Surg Asunto de la revista: ANGIOLOGIA / CARDIOLOGIA Año: 2020 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Prótesis Valvulares Cardíacas / Hemodinámica Límite: Humans Idioma: En Revista: Gen Thorac Cardiovasc Surg Asunto de la revista: ANGIOLOGIA / CARDIOLOGIA Año: 2020 Tipo del documento: Article País de afiliación: Estados Unidos
...