Test-retest repeatability of the pattern electroretinogram and flicker electroretinogram.
Resende, Arthur F; Sanvicente, Carina T; Eshraghi, Hamoon; Garcia, Alberto; Pickel, Kassandra; Zhang, Qiang; Waisbourd, Michael; Jay Katz, L.
Afiliación
  • Resende AF; Glaucoma Research Center, Wills Eye Hospital, 840 Walnut Street, Suite 1140, Philadelphia, PA, 19107, USA.
  • Sanvicente CT; Glaucoma Research Center, Wills Eye Hospital, 840 Walnut Street, Suite 1140, Philadelphia, PA, 19107, USA.
  • Eshraghi H; St. Luke's University Hospital, Bethlehem, PA, USA.
  • Garcia A; Diopsys Inc., Pine Brook, NJ, USA.
  • Pickel K; Diopsys Inc., Pine Brook, NJ, USA.
  • Zhang Q; Vickie and Jack Farber Vision Research Center, Wills Eye Hospital, Philadelphia, PA, USA.
  • Waisbourd M; Division of Ophthalmology, Sackler Faculty of Medicine, Tel-Aviv Medical Center, Tel-Aviv, Israel.
  • Jay Katz L; Glaucoma Research Center, Wills Eye Hospital, 840 Walnut Street, Suite 1140, Philadelphia, PA, 19107, USA. ljkatz@willseye.org.
Doc Ophthalmol ; 139(3): 185-195, 2019 12.
Article en En | MEDLINE | ID: mdl-31312944
ABSTRACT

PURPOSE:

To evaluate the repeatability of the steady-state pattern electroretinogram (PERG) and full-field flicker electroretinogram (Flicker ERG) protocols, delivered by the office-based Neuro Optic Vision Assessment (NOVA)™ testing platform, in healthy subjects.

METHODS:

Healthy individuals underwent PERG (16° and 24°) and Flicker ERG [fixed luminance (FL) and multi-luminance (ML)] testing protocols. Test-retest repeatability of protocols was calculated using intra-class correlation coefficients (ICC). Reference values of the parameters of the aforementioned tests were also calculated.

RESULTS:

The ICCs for the PERG parameters ranged from 0.793 to 0.911 (p < 0.001). The ICCs for the Flicker ERG parameters ranged from 0.968 to 0.994 (p < 0.001). A linear regression analysis was applied to assess the impact of age on ERG responses. Age had a significant impact on all PERG parameters (16° or 24°). The phase response of the FL Flicker ERG significantly decreased with age (ß = - 0.837, p ≤ 0.001). The FL Flicker ERG Magnitude was also impacted with a significant quadratic effect of age (ß = - 0.0047, p = 0.0004). Similarly, the Phase Area Under the Curve (Phase AUC) of the ML Flicker ERG significantly declined with age (ß = - 0.007, p = 0.009), and the impact on the Magnitude AUC was significant as well, with a negative quadratic age effect.

CONCLUSIONS:

The PERG and Flicker ERG protocols, delivered by an office-based testing platform, were shown to have good-to-excellent test-retest repeatability when tests were performed in the same order and in immediate succession.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Retina / Electrorretinografía Tipo de estudio: Observational_studies / Prevalence_studies / Prognostic_studies / Risk_factors_studies Límite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Doc Ophthalmol Año: 2019 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Retina / Electrorretinografía Tipo de estudio: Observational_studies / Prevalence_studies / Prognostic_studies / Risk_factors_studies Límite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Doc Ophthalmol Año: 2019 Tipo del documento: Article País de afiliación: Estados Unidos