Variation in target volume and centroid position due to breath holding during four-dimensional computed tomography scanning: A phantom study.
Miyamae, Yuta; Akimoto, Mami; Sasaki, Makoto; Fujimoto, Takahiro; Yano, Shinsuke; Nakamura, Mitsuhiro.
Afiliación
  • Miyamae Y; Division of Clinical Radiology Service, Kyoto University Hospital, Kyoto, Japan.
  • Akimoto M; Department of Radiological Technology, Radiological Diagnosis, National Cancer Center Hospital, Tokyo, Japan.
  • Sasaki M; Department of Radiation Oncology, Kurashiki Central Hospital, Okayama, Japan.
  • Fujimoto T; Division of Clinical Radiology Service, Kyoto University Hospital, Kyoto, Japan.
  • Yano S; Division of Clinical Radiology Service, Kyoto University Hospital, Kyoto, Japan.
  • Nakamura M; Division of Clinical Radiology Service, Kyoto University Hospital, Kyoto, Japan.
J Appl Clin Med Phys ; 21(1): 11-17, 2020 Jan.
Article en En | MEDLINE | ID: mdl-31385421
This study investigated the effects of respiratory motion, including unwanted breath holding, on the target volume and centroid position on four-dimensional computed tomography (4DCT) imaging. Cine 4DCT images were reconstructed based on a time-based sorting algorithm, and helical 4DCT images were reconstructed based on both the time-based sorting algorithm and an amplitude-based sorting algorithm. A spherical object 20 mm in diameter was moved according to several simulated respiratory motions, with a motion period of 4.0 s and maximum amplitude of 5 mm. The object was extracted automatically, and the target volume and centroid position in the craniocaudal direction were measured using a treatment planning system. When the respiratory motion included unwanted breath-holding times shorter than the breathing cycle, the root mean square errors (RSME) between the reference and imaged target volumes were 18.8%, 14.0%, and 5.5% in time-based images in cine mode, time-based images in helical mode, and amplitude-based images in helical mode, respectively. In helical mode, the RSME between the reference and imaged centroid position was reduced from 1.42 to 0.50 mm by changing the reconstruction method from time- to amplitude-based sorting. When the respiratory motion included unwanted breath-holding times equal to the breathing cycle, the RSME between the reference and imaged target volumes were 19.1%, 24.3%, and 15.6% in time-based images in cine mode, time-based images in helical mode, and amplitude-based images in helical mode, respectively. In helical mode, the RSME between the reference and imaged centroid position was reduced from 1.61 to 0.83 mm by changing the reconstruction method from time- to amplitude-based sorting. With respiratory motion including breath holding of shorter duration than the breathing cycle, the accuracies of the target volume and centroid position were improved by amplitude-based sorting, particularly in helical 4DCT.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Algoritmos / Planificación de la Radioterapia Asistida por Computador / Interpretación de Imagen Radiográfica Asistida por Computador / Fantasmas de Imagen / Tomografía Computarizada Cuatridimensional / Contencion de la Respiración / Neoplasias Pulmonares Límite: Humans Idioma: En Revista: J Appl Clin Med Phys Asunto de la revista: BIOFISICA Año: 2020 Tipo del documento: Article País de afiliación: Japón Pais de publicación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Algoritmos / Planificación de la Radioterapia Asistida por Computador / Interpretación de Imagen Radiográfica Asistida por Computador / Fantasmas de Imagen / Tomografía Computarizada Cuatridimensional / Contencion de la Respiración / Neoplasias Pulmonares Límite: Humans Idioma: En Revista: J Appl Clin Med Phys Asunto de la revista: BIOFISICA Año: 2020 Tipo del documento: Article País de afiliación: Japón Pais de publicación: Estados Unidos