ABSTRACT

PURPOSE:

ICRU Report 90 on Key Data for Ionizing-Radiation Dosimetry Measurement Standards and Applications (2014) has reaffirmed the recommended value of the mean energy required to create an ion pair in air, Wair , to be 33.97(12) eV. The report also indicates that this "constant" of radiation dosimetry is energy independent above 10 keV, since there is no theoretical or experimental evidence to the contrary. The goal of this investigation is to obtain additional experimental determinations of Wair in high energy beams and thus to verify the suggested energy independence.

METHODS:

Wair can be evaluated by combining ionometric and calorimetric measurements with a calculated ratio of the absorbed dose in the ion chamber air cavity and that of the calorimeter absorbing element. In this investigation, a graphite parallel plate chamber and a graphite calorimeter were used and the dose ratio was calculated using the EGSnrc Monte Carlo code. Measurements were made in electron beams from the NRC Vickers linear accelerator at two incident energies, 20 and 35 MeV. A range of average energies at the measurement point were obtained by inserting graphite plates in the primary beam.

RESULTS:

The average value of Wair obtained in this investigation is 33.85(18) eV which is consistent with the recommended value of 33.97(12) eV where the number in brackets represents the combined standard uncertainty of the value, referring to the corresponding last digits. The individual values of Wair do not show any statistically significant energy dependence.

CONCLUSION:

The overall combined uncertainty of 0.5% meets the original target of the investigation. A larger-scale investigation, involving more individual energy points and a wider range of electron energies is required to go further and, for example, comment on the Wair energy dependency question raised by Tessier et al. [Med. Phys. 2018;45370-381].
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Radiometría / Electrones / Grafito Tipo de estudio: Health_economic_evaluation Idioma: En Revista: Med Phys Año: 2019 Tipo del documento: Article País de afiliación: Canadá Pais de publicación: EEUU / ESTADOS UNIDOS / ESTADOS UNIDOS DA AMERICA / EUA / UNITED STATES / UNITED STATES OF AMERICA / US / USA

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Radiometría / Electrones / Grafito Tipo de estudio: Health_economic_evaluation Idioma: En Revista: Med Phys Año: 2019 Tipo del documento: Article País de afiliación: Canadá Pais de publicación: EEUU / ESTADOS UNIDOS / ESTADOS UNIDOS DA AMERICA / EUA / UNITED STATES / UNITED STATES OF AMERICA / US / USA