Subcutaneous implantable cardioverter defibrillators in patients with left ventricular assist devices: case report and comprehensive review.
López-Gil, María; Fontenla, Adolfo; Delgado, Juan F; Rodríguez-Muñoz, Daniel.
Afiliación
  • López-Gil M; Cardiac Electrophysiology & Arrhythmia Unit, Cardiology Department, 12 de Octubre University Hospital, Avda de Córdoba s/n, Madrid, Spain.
  • Fontenla A; I+12 Investigation Institute, 12 de Octubre University Hospital, Avda de Córdoba s/n, Madrid, Spain.
  • Delgado JF; Cardiac Electrophysiology & Arrhythmia Unit, Cardiology Department, 12 de Octubre University Hospital, Avda de Córdoba s/n, Madrid, Spain.
  • Rodríguez-Muñoz D; I+12 Investigation Institute, 12 de Octubre University Hospital, Avda de Córdoba s/n, Madrid, Spain.
Eur Heart J Case Rep ; 3(2)2019 Jun 01.
Article en En | MEDLINE | ID: mdl-31449611
BACKGROUND: Left ventricular assist devices (LVAD) are increasingly used in patients with advanced heart failure, many of whom have been or will be implanted with an implantable cardioverter defibrillator (ICD). Interaction between both devices is a matter of concern. Subcutaneous ICD (S-ICD) obtains its signals through subcutaneous vectors, which poses special challenges with regards to adequate performance following LVAD implantation. CASE SUMMARY: We describe the case of a 24-year-old man implanted with an S-ICD because of idiopathic dilated cardiomyopathy, severe biventricular dysfunction, and self-limiting sustained ventricular tachycardias. After the implantation of a HeartMate 3™ (Left Ventricular Assist System, Abbott) several months later, the S-ICD became useless because of inappropriate sensing due to electromagnetic interference and attenuation of QRS voltage. DISCUSSION: We reviewed the reported cases in PubMed about the concomitant use of S-ICD and LVAD. Seven case reports about the performance of S-ICD in patients with an LVAD were identified, with discordant results. From these articles, we analyse the potential causes for these differing results. Pump location and operating rates in LVAD, as well as changes in the subcutaneous-electrocardiogram detected by the S-ICD after LVAD implantation are related to sensing disturbances when used in the same patient.