A Novel Quadripolar Active Fixation Left-Ventricular Pacing Lead for Cardiac Resynchronization Therapy: Initial United Kingdom Experience.
Chapman, Michael; Bates, Matthew G D; Behar, Jonathan M; Williams, Ian; Dewhurst, Matthew; Monkhouse, Christopher; Hayward, Carl; Muthumala, Amal; Chow, Anthony; Linker, Nicholas J; Thornley, Andrew R; Turley, Andrew J.
Afiliación
  • Chapman M; James Cook University Hospital, Middlesbrough, United Kingdom. Electronic address: michael.chapman6@nhs.net.
  • Bates MGD; James Cook University Hospital, Middlesbrough, United Kingdom.
  • Behar JM; St. Bartholomew's Hospital, United Kingdom.
  • Williams I; Norfolk and Norwich University Hospital, United Kingdom.
  • Dewhurst M; James Cook University Hospital, Middlesbrough, United Kingdom.
  • Monkhouse C; St. Bartholomew's Hospital, United Kingdom.
  • Hayward C; St. Bartholomew's Hospital, United Kingdom.
  • Muthumala A; St. Bartholomew's Hospital, United Kingdom.
  • Chow A; St. Bartholomew's Hospital, United Kingdom.
  • Linker NJ; James Cook University Hospital, Middlesbrough, United Kingdom.
  • Thornley AR; James Cook University Hospital, Middlesbrough, United Kingdom.
  • Turley AJ; James Cook University Hospital, Middlesbrough, United Kingdom.
JACC Clin Electrophysiol ; 5(9): 1028-1035, 2019 09.
Article en En | MEDLINE | ID: mdl-31537331
ABSTRACT

OBJECTIVES:

This study sought to assess immediate and short-term performance of the Medtronic Attain Stability Quadripolar 4798 lead (Medtronic, Dublin, Ireland).

BACKGROUND:

Cardiac resynchronization therapy (CRT) is an established treatment for appropriately selected patients with left ventricular (LV) systolic dysfunction. The most common reason for failure to implant a lead is the lack of a suitable epicardial vein, due either to an absent vessel in the target site, an unacceptably high threshold, lead instability, phrenic nerve stimulation, or a combination of reasons. In August 2017, a novel quadripolar active fixation LV lead (Medtronic) was released. This paper reports the initial clinical experience with lead implantation and specifically immediate and short-term pacing parameters across 3 United Kingdom centers.

METHODS:

Consecutive patients eligible for CRT were deemed suitable for this lead. Immediate and short-term lead performance data regarding LV threshold, impedance, and displacement rates were collected at standard pacing checks (1 day, 5 weeks, 3 months, and 9 months post-implantation).

RESULTS:

CRT using this lead was attempted in 82 cases and was successful in 81 cases (98.8%). LV thresholds and impedance levels were 1.22 ± 0.75 V and 737 ± 319 Ω at implantation; 1.16 ± 0.71 V and 597 ± 218 Ω at day 1; 1.02 ± 0.48 V and 579 ± 148 Ω at week 6; 0.98 ± 0.49 V and 569 ± 133 Ω at 3 months; and 1.06 ± 0.48 V and 570 ± 140 Ω at 9 months. As of the publication of this paper, no LV lead has been displaced.

CONCLUSIONS:

CRT using the Medtronic lead was successful in more than 98% of the patients. Short-to-medium-term data regarding lead performance and stability were excellent, with zero displacements as of the publication of this paper.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Terapia de Resincronización Cardíaca / Dispositivos de Terapia de Resincronización Cardíaca Tipo de estudio: Observational_studies Límite: Aged / Aged80 / Female / Humans / Male / Middle aged País/Región como asunto: Europa Idioma: En Revista: JACC Clin Electrophysiol Año: 2019 Tipo del documento: Article

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Terapia de Resincronización Cardíaca / Dispositivos de Terapia de Resincronización Cardíaca Tipo de estudio: Observational_studies Límite: Aged / Aged80 / Female / Humans / Male / Middle aged País/Región como asunto: Europa Idioma: En Revista: JACC Clin Electrophysiol Año: 2019 Tipo del documento: Article
...