Neonatal endotracheal flowmeter for tidal volume, airway pressure, and end-tidal gas.
J Appl Physiol (1985) ; 58(3): 1023-5, 1985 Mar.
Article en En | MEDLINE | ID: mdl-3156842
ABSTRACT
We have designed a new endotracheal flowmeter to measure tidal volume, phasic and mean airway pressures, inspiratory time, and end-tidal PCO2 and PO2 in intubated infants. The flowmeter is light (11 g) and adds minimal dead space (1.0 ml) and resistance (2 cmH2O X 100 ml- X s) to the infant's airway. The volume signal (less than or equal to 10 ml) is linear to 7 Hz, and end-tidal gases can be measured at respiratory rates of 90 breaths/min. This flowmeter is particularly valuable for evaluation of rapid mechanical ventilation of very low birth weight infants.
Asunto(s)
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Reología / Recién Nacido de Bajo Peso / Mediciones del Volumen Pulmonar Límite: Humans / Newborn Idioma: En Revista: J Appl Physiol (1985) Asunto de la revista: FISIOLOGIA Año: 1985 Tipo del documento: Article Pais de publicación: EEUU / ESTADOS UNIDOS / ESTADOS UNIDOS DA AMERICA / EUA / UNITED STATES / UNITED STATES OF AMERICA / US / USA
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Reología / Recién Nacido de Bajo Peso / Mediciones del Volumen Pulmonar Límite: Humans / Newborn Idioma: En Revista: J Appl Physiol (1985) Asunto de la revista: FISIOLOGIA Año: 1985 Tipo del documento: Article Pais de publicación: EEUU / ESTADOS UNIDOS / ESTADOS UNIDOS DA AMERICA / EUA / UNITED STATES / UNITED STATES OF AMERICA / US / USA