A New Shielding Curtain for Protection of Intraoperative Radiation During Minimally Invasive Spine Surgery.
Park, Se Hoon; Ko, Myeong Jin; Park, Yong Sook; Yun, Jaehan; Byun, Hongsik; Park, Seung Won.
Afiliación
  • Park SH; Department of Neurosurgery, Chung-Ang University Hospital, Seoul, Korea.
  • Ko MJ; Department of Neurosurgery, Chung-Ang University Hospital, Seoul, Korea.
  • Park YS; Department of Neurosurgery, Chung-Ang University Hospital, Seoul, Korea.
  • Yun J; Department of Chemical Engineering, Keimyung University, Daegu, Korea.
  • Byun H; Department of Chemical Engineering, Keimyung University, Daegu, Korea.
  • Park SW; Department of Neurosurgery, Chung-Ang University Hospital, Seoul, Korea.
Neurospine ; 17(1): 288-293, 2020 Mar.
Article en En | MEDLINE | ID: mdl-31614404
ABSTRACT

OBJECTIVE:

Gradually increasing number of minimally invasive spine surgery (MISS), there is an increasing risk of radiation exposure to medical personnel during the surgery. We measured the radiation exposure of the operating room personnel during MISS, tried to find the riskiest person, and checked the effectiveness of a new lead-composite shielding curtain.

METHODS:

Radiation exposure of medical staffs (operator, first assistant, anesthesiologist, and scrub nurse) involved in MISS procedures of 35 patients without shielding curtain (nonshield group) and 35 patients with shielding curtain (shield group). The shielding curtain had 0.25-mm nominal lead equivalent and was mounted on 2 frame bars fixed on the operating table.

RESULTS:

In the nonshield group, radiation exposure was significantly higher in the order of operator > first assistant > scrub nurse > anesthesiologist (p < 0.001) during both anteroposterior (AP) and lateral views. In the shield group, the radiation exposure of the operator and the scrub nurse decreased significantly by 94.1% and 76.4% in AP view (p < 0.001), and by 96.3% and 73.9% in lateral view (p < 0.001), respectively.

CONCLUSION:

Since the radiation dose of the operator was highest in a C-arm-guided MISS, there is a high priority need to protect the operator from the radiation exposure. The shielding curtain could most effectively reduce the radiation exposure of the operator.