Muluhngwi, Penn; Sharp, Cierra N; Pozzi, Nicole; Elin, Ronald J; Jortani, Saeed A.
Afiliación
  • Muluhngwi P; Department of Pathology and Laboratory Medicine, University of Louisville, Louisville, KY.
  • Sharp CN; Department of Pathology and Laboratory Medicine, University of Louisville, Louisville, KY.
  • Pozzi N; University of Louisville Hospital, Louisville, KY.
  • Elin RJ; Department of Pathology and Laboratory Medicine, University of Louisville, Louisville, KY.
  • Jortani SA; Department of Pathology and Laboratory Medicine, University of Louisville, Louisville, KY; saeed.jortani@louisville.edu.
J Appl Lab Med ; 4(3): 323-330, 2019 11.
Article en En | MEDLINE | ID: mdl-31659070
ABSTRACT

BACKGROUND:

κ and λ free light chains (FLCs) are monitored to aid in the diagnosis of plasma cell disorders. Our goal was to validate the Diazyme Human κ and λ assays on Beckman Coulter UniCel DxC 800 Synchron and compare to Freelite κ and λ assays on Roche Cobas Integra.

METHODS:

Linearity verification, within- and between-run precision, method comparison, and reference range (RR) verification were conducted using CLSI guidelines. Statistical analysis was performed using EP Evaluator®. Mean, SD, CV, and bias were determined.

RESULTS:

Diazyme κ FLC assay was linear within 0.00-191.00 mg/L. Diazyme λ FLC assay was linear within 0.00-205.30 mg/L. Diazyme κ FLC QC1 had a mean of 16.70 mg/L, CV of 7.0%. QC2 had a mean of 33.37 mg/L, CV of 2.6%. Diazyme λ FLC QC1 had a mean of 21.73 mg/L, CV of 2.3%. QC2 had a mean of 42.05 mg/L, CV of 1.5%. Bias of DxC-Diazyme FLCs compared to Integra-Freelite FLCs was -2.55 mg/L (κ FLC), and 4.54 mg/L (λ FLC). Qualitative comparison of κ FLC assays indicated 100% agreement for both normal and abnormal values. For λ FLC assay, agreement was 95% for normal values and 75% for abnormal values. For κ/λ ratio there was 50% agreement for normal values, and 100% for abnormal values. For RR verification, 1 sample was outside the Diazyme κ RR. For λ, all samples were within the manufacturer's RR.

CONCLUSIONS:

Diazyme assays for FLCs have excellent precision and accuracy and are comparable to Freelite assays.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Pruebas Inmunológicas / Cadenas kappa de Inmunoglobulina / Cadenas lambda de Inmunoglobulina / Aprobación de Pruebas de Diagnóstico Tipo de estudio: Diagnostic_studies / Guideline / Qualitative_research Límite: Humans País/Región como asunto: America do norte Idioma: En Revista: J Appl Lab Med Año: 2019 Tipo del documento: Article

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Pruebas Inmunológicas / Cadenas kappa de Inmunoglobulina / Cadenas lambda de Inmunoglobulina / Aprobación de Pruebas de Diagnóstico Tipo de estudio: Diagnostic_studies / Guideline / Qualitative_research Límite: Humans País/Región como asunto: America do norte Idioma: En Revista: J Appl Lab Med Año: 2019 Tipo del documento: Article