Your browser doesn't support javascript.
loading
Effect of Optimum Torque Reverse Motion on Torque and Force Generation during Root Canal Instrumentation with Crown-down and Single-length Techniques.
Kimura, Shunsuke; Ebihara, Arata; Maki, Keiichiro; Nishijo, Miki; Tokita, Daisuke; Okiji, Takashi.
Afiliación
  • Kimura S; Department of Pulp Biology and Endodontics, Division of Oral Health Sciences, Graduate School of Medical and Dental Sciences, Tokyo Medical and Dental University (TMDU), Tokyo, Japan.
  • Ebihara A; Department of Pulp Biology and Endodontics, Division of Oral Health Sciences, Graduate School of Medical and Dental Sciences, Tokyo Medical and Dental University (TMDU), Tokyo, Japan. Electronic address: a.ebihara.endo@tmd.ac.jp.
  • Maki K; Department of Pulp Biology and Endodontics, Division of Oral Health Sciences, Graduate School of Medical and Dental Sciences, Tokyo Medical and Dental University (TMDU), Tokyo, Japan.
  • Nishijo M; Department of Pulp Biology and Endodontics, Division of Oral Health Sciences, Graduate School of Medical and Dental Sciences, Tokyo Medical and Dental University (TMDU), Tokyo, Japan.
  • Tokita D; Department of Pulp Biology and Endodontics, Division of Oral Health Sciences, Graduate School of Medical and Dental Sciences, Tokyo Medical and Dental University (TMDU), Tokyo, Japan.
  • Okiji T; Department of Pulp Biology and Endodontics, Division of Oral Health Sciences, Graduate School of Medical and Dental Sciences, Tokyo Medical and Dental University (TMDU), Tokyo, Japan.
J Endod ; 46(2): 232-237, 2020 Feb.
Article en En | MEDLINE | ID: mdl-31889584
ABSTRACT

INTRODUCTION:

Optimum torque reverse (OTR) motion is a torque-sensitive reciprocal motion in which the motor rotates in alternating 90° counterclockwise and 180° clockwise rotation when the torque exceeds a predetermined value. This study aimed to examine whether OTR motion contributes to torque and force reduction during nickel-titanium rotary instrumentation with the crown-down or single-length technique.

METHODS:

Twenty-eight simulated straight canals in resin blocks were divided into 2 groups according to the type of motion (OTR or continuous rotation). The groups were further subdivided according to the preparation technique (crown-down or single-length technique, n = 7 each). Automated root canal instrumentation was performed with a torque/force analyzing device (300 rpm, up-and-down speed of 10 mm/min) and EndoWave instruments (FKG Dentaire, La-Chaux-de-Fonds, Switzerland) to size #25/0.06 taper. Maximum torque and apical force were recorded and analyzed with analysis of variance and the Bonferroni test.

RESULTS:

During the crown-down preparation phase (#35/0.08, #30/0.06, #25/0.06, and #20/0.06), OTR motion developed lower maximum torque and upward force (representing the screw-in force) than continuous rotation. During the apical preparation phase (#25/0.06), OTR motion generated significantly lower maximum clockwise and counterclockwise torque (P < .05) when the single-length technique was used and significantly lower maximum upward force regardless of the preparation technique (P < .05) compared with continuous rotation.

CONCLUSIONS:

Under the present experimental conditions, OTR motion reduced both torque and screw-in force during the crown-down preparation phase of the crown-down technique and during the apical preparation phase of the single-length technique.