CdTe compact gamma camera for coded aperture imaging in radioguided surgery.
Russo, Paolo; Di Lillo, Francesca; Corvino, Veronica; Frallicciardi, Paola Maria; Sarno, Antonio; Mettivier, Giovanni.
Afiliación
  • Russo P; Universita' di Napoli Federico II, Dipartimento di Fisica "Ettore Pancini", I-80126 Napoli, Italy; INFN Sezione di Napoli, I-80126 Napoli, Italy.
  • Di Lillo F; Universita' di Napoli Federico II, Dipartimento di Fisica "Ettore Pancini", I-80126 Napoli, Italy; INFN Sezione di Napoli, I-80126 Napoli, Italy; Elettra-Sincrotrone Trieste S.C.p.A., Trieste, Italy.
  • Corvino V; Universita' di Napoli Federico II, Dipartimento di Fisica "Ettore Pancini", I-80126 Napoli, Italy.
  • Frallicciardi PM; Universita' di Napoli Federico II, Dipartimento di Fisica "Ettore Pancini", I-80126 Napoli, Italy; INFN Sezione di Napoli, I-80126 Napoli, Italy; Azienda Ospedaliera Universitaria, Ospedali Riuniti, Foggia, Italy.
  • Sarno A; Universita' di Napoli Federico II, Dipartimento di Fisica "Ettore Pancini", I-80126 Napoli, Italy; INFN Sezione di Napoli, I-80126 Napoli, Italy.
  • Mettivier G; Universita' di Napoli Federico II, Dipartimento di Fisica "Ettore Pancini", I-80126 Napoli, Italy; INFN Sezione di Napoli, I-80126 Napoli, Italy. Electronic address: mettivier@na.infn.it.
Phys Med ; 69: 223-232, 2020 Jan.
Article en En | MEDLINE | ID: mdl-31918374
The aim of this work was to assess the performance of a prototype compact gamma camera (MediPROBE) based on a CdTe semiconductor hybrid pixel detector, for coded aperture imaging. This probe can be adopted for various tasks in nuclear medicine such as preoperative sentinel lymph node localization, breast imaging with 99mTc radiotracers and thyroid imaging, and in general in radioguided surgery tasks. The hybrid detector is an assembly of a 1-mm thick CdTe semiconductor detector bump-bonded to a photon-counting CMOS readout circuit of the Medipix2 series or energy-sensitive Timepix detector. MediPROBE was equipped with a set of two coded aperture masks with 0.07-mm or 0.08-mm diameter holes. We performed laboratory measurements of field of view, system spatial resolution, and signal-difference-to-noise ratio, by using gamma-emitting radioactive sources (109Cd, 125I, 241Am, 99mTc). The system spatial resolution in the lateral direction was 0.56 mm FWHM (coded aperture mask with holes of 0.08 mm and a 60 keV source) at a source-collimator distance of 50 mm and a field of view of 40 mm by side. Correspondingly, the longitudinal resolution in 3D source localization tasks was about 3 mm. MediPROBE showed a significant improvement in terms of spatial resolution when equipped with the high-resolution coded apertures, with respect to the performance previously reported with 1-2 mm pinhole apertures as well as with respect to adopting a 0.35 mm pinhole aperture.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Cámaras gamma / Radiocirugia Límite: Humans Idioma: En Revista: Phys Med Asunto de la revista: BIOFISICA / BIOLOGIA / MEDICINA Año: 2020 Tipo del documento: Article País de afiliación: Italia Pais de publicación: Italia

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Cámaras gamma / Radiocirugia Límite: Humans Idioma: En Revista: Phys Med Asunto de la revista: BIOFISICA / BIOLOGIA / MEDICINA Año: 2020 Tipo del documento: Article País de afiliación: Italia Pais de publicación: Italia