Benchmarking a commercial proton therapy solution: The Paul Scherrer Institut experience.
Rosas, Sara; Belosi, Francesca M; Bizzocchi, Nicola; Böhlen, Till; Zepter, Stefan; Morach, Petra; Lomax, Antony J; Weber, Damien C; Hrbacek, Jan.
Afiliación
  • Rosas S; Zentrum für Protonentherapie, Paul Scherrer Institut, Villigen, Switzerland.
  • Belosi FM; Zentrum für Protonentherapie, Paul Scherrer Institut, Villigen, Switzerland.
Br J Radiol ; 93(1107): 20190920, 2020 Mar.
Article en En | MEDLINE | ID: mdl-31944827
OBJECTIVE: For the past 20 years, Paul Scherrer Institut (PSI) has treated more than 1500 patients with deep-seated tumors using PSI-Plan, an in-house developed treatment planning system (TPS) used for proton beam scanning proton therapy, in combination with its home-built gantries. The goal of the present work is to benchmark the performance of a new TPS/Gantry system for proton therapy centers which have established already a baseline standard of care. METHODS AND MATERIALS: A total of 31 cases (=52 plans) distributed around 7 anatomical sites and 12 indications were randomly selected and re-planned using Eclipse™. The resulting plans were compared with plans formerly optimized in PSI-Plan, in terms of target coverage, plan quality, organ-at-risk (OAR) sparing and number of delivered pencil beams. RESULTS: Our results show an improvement on target coverage and homogeneity when using Eclipse™ while PSI-Plan showed superior plan conformity. As for OAR sparing, both TPS achieved the clinical constraints. The number of pencil beams required per plan was on average 3.4 times higher for PSI-Plan. CONCLUSION: Both systems showed a good capacity to produce satisfactory plans, with Eclipse™ being able to achieve better target coverage and plan homogeneity without compromising OARs. ADVANCES IN KNOWLEDGE: A benchmark between a clinically tested and validated system with a commercial solution is of interest for emerging proton therapy, equipped with commercial systems and no previous experience with proton beam scanning.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Planificación de la Radioterapia Asistida por Computador / Instituciones Oncológicas / Benchmarking / Mejoramiento de la Calidad / Terapia de Protones / Neoplasias Límite: Humans País/Región como asunto: Europa Idioma: En Revista: Br J Radiol Año: 2020 Tipo del documento: Article País de afiliación: Suiza Pais de publicación: Reino Unido

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Planificación de la Radioterapia Asistida por Computador / Instituciones Oncológicas / Benchmarking / Mejoramiento de la Calidad / Terapia de Protones / Neoplasias Límite: Humans País/Región como asunto: Europa Idioma: En Revista: Br J Radiol Año: 2020 Tipo del documento: Article País de afiliación: Suiza Pais de publicación: Reino Unido