Your browser doesn't support javascript.
loading
Rapid whole-brain quantitative magnetization transfer imaging using 3D selective inversion recovery sequences.
Cronin, Matthew J; Xu, Junzhong; Bagnato, Francesca; Gochberg, Daniel F; Gore, John C; Dortch, Richard D.
Afiliación
  • Cronin MJ; Vanderbilt University Medical Center, Department Radiology and Radiological Sciences, Nashville, TN, United States of America; Vanderbilt University Medical Center, Institute of Imaging Science, Nashville, TN, United States of America.
  • Xu J; Vanderbilt University Medical Center, Department Radiology and Radiological Sciences, Nashville, TN, United States of America; Vanderbilt University Medical Center, Institute of Imaging Science, Nashville, TN, United States of America; Vanderbilt University, Department of Physics and Astronomy, Nash
  • Bagnato F; Vanderbilt University Medical Center, Department of Neurology, Neuro-Immunology Division/Neuro-Imaging Unit, Nashville, TN, United States of America.
  • Gochberg DF; Vanderbilt University Medical Center, Department Radiology and Radiological Sciences, Nashville, TN, United States of America; Vanderbilt University Medical Center, Institute of Imaging Science, Nashville, TN, United States of America; Vanderbilt University, Department of Physics and Astronomy, Nash
  • Gore JC; Vanderbilt University Medical Center, Department Radiology and Radiological Sciences, Nashville, TN, United States of America; Vanderbilt University Medical Center, Institute of Imaging Science, Nashville, TN, United States of America; Vanderbilt University, Department of Physics and Astronomy, Nash
  • Dortch RD; Vanderbilt University Medical Center, Department Radiology and Radiological Sciences, Nashville, TN, United States of America; Vanderbilt University Medical Center, Institute of Imaging Science, Nashville, TN, United States of America; Vanderbilt University, Department of Biomedical Engineering, Nas
Magn Reson Imaging ; 68: 66-74, 2020 05.
Article en En | MEDLINE | ID: mdl-32004710
ABSTRACT
Selective inversion recovery (SIR) is a quantitative magnetization transfer (qMT) method that provides estimates of parameters related to myelin content in white matter, namely the macromolecular pool-size-ratio (PSR) and the spin-lattice relaxation rate of the free pool (R1f), without the need for independent estimates of ∆B0, B1+, and T1. Although the feasibility of performing SIR in the human brain has been demonstrated, the scan times reported previously were too long for whole-brain applications. In this work, we combined optimized, short-TR acquisitions, SENSE/partial-Fourier accelerations, and efficient 3D readouts (turbo spin-echo, SIR-TSE; echo-planar imaging, SIR-EPI; and turbo field echo, SIR-TFE) to obtain whole-brain data in 18, 10, and 7 min for SIR-TSE, SIR-EPI, SIR-TFE, respectively. Based on numerical simulations, all schemes provided accurate parameter estimates in large, homogenous regions; however, the shorter SIR-TFE scans underestimated focal changes in smaller lesions due to blurring. Experimental studies in healthy subjects (n = 8) yielded parameters that were consistent with literature values and repeatable across scans (coefficient of variation PSR = 2.2-6.4%, R1f = 0.6-1.4%) for all readouts. Overall, SIR-TFE parameters exhibited the lowest variability, while SIR-EPI parameters were adversely affected by susceptibility-related image distortions. In patients with relapsing remitting multiple sclerosis (n = 2), focal changes in SIR parameters were observed in lesions using all three readouts; however, contrast was reduced in smaller lesions for SIR-TFE, which was consistent with the numerical simulations. Together, these findings demonstrate that efficient, accurate, and repeatable whole-brain SIR can be performed using 3D TFE, EPI, or TSE readouts; however, the appropriate readout should be tailored to the application.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Encéfalo / Mapeo Encefálico / Sustancia Blanca Tipo de estudio: Prognostic_studies Límite: Adult / Female / Humans / Male Idioma: En Revista: Magn Reson Imaging Año: 2020 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Encéfalo / Mapeo Encefálico / Sustancia Blanca Tipo de estudio: Prognostic_studies Límite: Adult / Female / Humans / Male Idioma: En Revista: Magn Reson Imaging Año: 2020 Tipo del documento: Article País de afiliación: Estados Unidos