Free-running cardiac magnetic resonance fingerprinting: Joint T1/T2 map and Cine imaging.
Jaubert, O; Cruz, G; Bustin, A; Schneider, T; Koken, P; Doneva, M; Rueckert, D; Botnar, R M; Prieto, C.
Afiliación
  • Jaubert O; School of Biomedical Engineering and Imaging Sciences, King's College London, London, United Kingdom. Electronic address: olivier.jaubert@kcl.ac.uk.
  • Cruz G; School of Biomedical Engineering and Imaging Sciences, King's College London, London, United Kingdom.
  • Bustin A; School of Biomedical Engineering and Imaging Sciences, King's College London, London, United Kingdom.
  • Schneider T; Philips Healthcare, Guilford, United Kingdom.
  • Koken P; Philips Research Europe, Hamburg, Germany.
  • Doneva M; Philips Research Europe, Hamburg, Germany.
  • Rueckert D; Department of Computing, Imperial College London, London, United Kingdom.
  • Botnar RM; School of Biomedical Engineering and Imaging Sciences, King's College London, London, United Kingdom.
  • Prieto C; School of Biomedical Engineering and Imaging Sciences, King's College London, London, United Kingdom; Escuela de Ingeniería, Pontificia Universidad Católica de Chile, Santiago, Chile.
Magn Reson Imaging ; 68: 173-182, 2020 05.
Article en En | MEDLINE | ID: mdl-32061964
ABSTRACT

PURPOSE:

To develop and evaluate a novel non-ECG triggered 2D magnetic resonance fingerprinting (MRF) sequence allowing for simultaneous myocardial T1 and T2 mapping and cardiac Cine imaging.

METHODS:

Cardiac MRF (cMRF) has been recently proposed to provide joint T1/T2 myocardial mapping by triggering the acquisition to mid-diastole and relying on a subject-dependent dictionary of MR signal evolutions to generate the maps. In this work, we propose a novel "free-running" (non-ECG triggered) cMRF framework for simultaneous myocardial T1 and T2 mapping and cardiac Cine imaging in a single scan. Free-running cMRF is based on a transient state bSSFP acquisition with tiny golden angle radial readouts, varying flip angle and multiple adiabatic inversion pulses. The acquired data is retrospectively gated into several cardiac phases, which are reconstructed with an approach that combines parallel imaging, low rank modelling and patch-based high-order tensor regularization. Free-running cMRF was evaluated in a standardized phantom and ten healthy subjects. Comparison with reference spin-echo, MOLLI, SASHA, T2-GRASE and Cine was performed.

RESULTS:

T1 and T2 values obtained with the proposed approach were in good agreement with reference phantom values (ICC(A,1) > 0.99). Reported values for myocardium septum T1 were 1043 ± 48 ms, 1150 ± 100 ms and 1160 ± 79 ms for MOLLI, SASHA and free-running cMRF respectively and for T2 of 51.7 ± 4.1 ms and 44.6 ± 4.1 ms for T2-GRASE and free-running cMRF respectively. Good agreement was observed between free-running cMRF and conventional Cine 2D ejection fraction (bias = -0.83%).

CONCLUSION:

The proposed free-running cardiac MRF approach allows for simultaneous assessment of myocardial T1 and T2 and Cine imaging in a single scan.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Espectroscopía de Resonancia Magnética / Imagen por Resonancia Cinemagnética / Electrocardiografía / Corazón / Miocardio Tipo de estudio: Observational_studies / Prognostic_studies Límite: Adult / Female / Humans / Male Idioma: En Revista: Magn Reson Imaging Año: 2020 Tipo del documento: Article Pais de publicación: HOLANDA / HOLLAND / NETHERLANDS / NL / PAISES BAJOS / THE NETHERLANDS

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Espectroscopía de Resonancia Magnética / Imagen por Resonancia Cinemagnética / Electrocardiografía / Corazón / Miocardio Tipo de estudio: Observational_studies / Prognostic_studies Límite: Adult / Female / Humans / Male Idioma: En Revista: Magn Reson Imaging Año: 2020 Tipo del documento: Article Pais de publicación: HOLANDA / HOLLAND / NETHERLANDS / NL / PAISES BAJOS / THE NETHERLANDS