Comparison of broad-range polymerase chain reaction and metagenomic next-generation sequencing for the diagnosis of prosthetic joint infection.
Wang, Chao-Xin; Huang, Zida; Fang, Xinyu; Li, Wenbo; Yang, Bin; Zhang, Wenming.
Afiliación
  • Wang CX; Department of Orthopedic Surgery, The First Affiliated Hospital of Fujian Medical University, Fuzhou 350004, China.
  • Huang Z; Department of Orthopedic Surgery, The First Affiliated Hospital of Fujian Medical University, Fuzhou 350004, China.
  • Fang X; Department of Orthopedic Surgery, The First Affiliated Hospital of Fujian Medical University, Fuzhou 350004, China.
  • Li W; Department of Orthopedic Surgery, The First Affiliated Hospital of Fujian Medical University, Fuzhou 350004, China.
  • Yang B; Department of Laboratory Medicine, The First Affiliated Hospital of Fujian Medical University, Fuzhou 350004, China.
  • Zhang W; Department of Orthopedic Surgery, The First Affiliated Hospital of Fujian Medical University, Fuzhou 350004, China. Electronic address: zhangwm0591@fjmu.edu.cn.
Int J Infect Dis ; 95: 8-12, 2020 Jun.
Article en En | MEDLINE | ID: mdl-32251799
ABSTRACT

PURPOSE:

The aims of our study were to (1) evaluate the concordance of both methods for detecting prosthetic joint infection (PJI) pathogens in joint fluid and to (2) clarify whether broad-range polymerase chain reaction (BR-PCR) can be used as a verification method for metagenomic next-generation sequencing (mNGS) for PJI diagnosis.

METHODS:

In total, 63 patients underwent total joint arthroplasty, with 45 PJI and 18 aseptic failure patients included. Joint fluids were sampled after antibiotics were withheld for more than 2 weeks, and then, culture, BR-PCR and mNGS were performed for all samples.

RESULTS:

The joint fluid BR-PCR sensitivity was 82.2%, which was not significantly different from that of mNGS (95.6%) or culture (77.8%). The specificities of the 3 methods were all 94.4%. BR-PCR failed to identify the pathogens in 1 polymicrobial infection patient and 4 fungal infection patients.

CONCLUSION:

mNGS was more sensitive than BR-PCR for detecting PJI pathogens in joint fluid. BR-PCR is insufficient for use as an mNGS verification method.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Reacción en Cadena de la Polimerasa / Infecciones Relacionadas con Prótesis / Secuenciación de Nucleótidos de Alto Rendimiento Tipo de estudio: Diagnostic_studies / Prognostic_studies Límite: Humans Idioma: En Revista: Int J Infect Dis Asunto de la revista: DOENCAS TRANSMISSIVEIS Año: 2020 Tipo del documento: Article País de afiliación: China

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Reacción en Cadena de la Polimerasa / Infecciones Relacionadas con Prótesis / Secuenciación de Nucleótidos de Alto Rendimiento Tipo de estudio: Diagnostic_studies / Prognostic_studies Límite: Humans Idioma: En Revista: Int J Infect Dis Asunto de la revista: DOENCAS TRANSMISSIVEIS Año: 2020 Tipo del documento: Article País de afiliación: China